- Series 6100 - Computerized Multi-Gas Calibration System

From
0.511.522.533.544.55 (0 votes)
Environics - Series 6100 - Computerized Multi-Gas Calibration System
The Environics Series 6100 Computerized Multi-Gas Calibration System is a computer controlled, state-of-the-art instrument for dynamic multi-point gas calibration of ambient air analyzers. The Series 6100 automatically performs zero, precision, span and multi-point calibrations using NO, NO2, SO2, CO, 03, hydrocarbons and other gases of interest. The 6100 meets all U.S. Environmental Protection Agency requirements. The Series 6100 consists of a single chassis supporting 2 thermal mass flow controllers, an ozone generation module, a mixing zone, a reaction chamber for gas phase titration, and control electronics.

PRODUCT FEATURES AND BENEFITS

  • User-friendly interactive software with plain language prompts is simple to use, reducing technician training time and virtually eliminating error.
  • Automatic calculation of dilution and span gas flows based on commanded concentration eliminates the need for manual computation and allows rapid transition from point to point.
  • Internally-stored mass flow controller calibration data improves accuracy by as much as a factor of ten and simplifies field recalibration. Mass flow controllers are factory calibrated at 11 points.
  • Internally-stored ozone generator calibration data insures linear, repeatable ozone generation without photometer control. Ozone generator performance meets U.S. EPA criteria for ozone transfer standards.
  • Ozone generator pressure compensation ensures repeatable and stable ozone generation at pressures other than the original calibration pressure.
  • Full, two-way RS-232 Serial Data Interface (standard) permits remote operation of the instrument including data collection, gas selection, initiation of zero, span and multi-point calibrations.
  • SOFTWARE
  • Concentration Mode: In response to software prompting, user selects gas port, span (cylinder) gas concentration, output gas flow (total) and output gas concentration. Series 6100 automatically delivers concentrations at the total flow specified.
  • Generate Ozone: Allows user to specify, then generate a precise concentration of ozone.
  • Gas Phase Titration: Utilizes blend and generate ozone routines to lead user through GPT using 'excess NO' method.
  • Flow Mode: Allows user to manually command a desired rate of flow for each mass flow controller, individually or together, and with or without ozone.
  • Display: Allows user to monitor flow rates for each mass flow controller separately. Also provides ozone oven block temperature during generate ozone and gas phase titration routines.
  • Maintain Ports: User enters the name of the span gas in the source cylinder, its concentration (ppm) and the port to which the cylinder is connected.

SPECIFICATIONS

Performance*

Accuracy
Concentration: ± 1.0% setpoint
Flow: ± 1.0% setpoint
Flow Repeatability: ± 0.05% F.S.O.

Warm up time: 30 minutes

* Performance specifications are valid when all Mass Flow Controllers are operating between 10% and 100% of full scale flow.
Mass flow controllers are calibrated using a NIST traceable Primary Flow Standard, using a Reference Temperature of 25°C (77°F) and a Reference Pressure of 760 mm Hg (29.92 in. Hg)

Ozone Generator
Concentration Range: 0.02 - 0.5 ppm at 5 - 10 slpm
Optional Ranges: 0.05 - 1.0 ppm at 5 - 10 slpm

Accuracy: /- 2% or /- 0.003 ppm
Repeatability: /- 2% or /- 0.003 ppm (for 30 days)

Mechanical

Inlets Balance: External 1/4' Swagelok*
Span(s): External 1/4' Swagelok*

Outlet
One external 1/4' Swagelok*
*(or compatible fitting)

Operating Pressures at inlets (flow dependent)
Minimum: 20 psig (1.38 Bar)
Nominal: 25 psig (1.72 Bar)
Maximum: 30 psig (2.07 Bar)

Wetted Surfaces
Tubing: Teflon®
MFCs: Stainless Steel
Seals: Viton®
Glass Chambers: Pyrex®

Operating temperatures
32° - 122° F (0° - 50° C)

Performance Temperature Range
59° - 95° F (15° - 35° C)

Weight
Standard: 20 lbs.
Including Options: 30 lbs.

Dimensions (w x h x d)
Portable: 17' x 7' x 15'
Rack: 19' x 7' x 15'

Electrical

Voltage: 100 to 250 VAC, 50/60 Hz
Current: 2 Amps (maximum)

Operating Modes

Front panel keypad
Internal timer control
RS-232 serial data interface
I/O control (8 inputs / 8 outputs)

Data I/O

RS-232 serial data interface
I/O Control (8 inputs / 8 outputs)

OPTIONS

  • Rack Mount
  • Extra Gas Inlets
  • Solenoid Valve on Output
  • Permeation Oven
  • 3rd Mass Flow Controller
Environics, Inc.
Environics, Inc.
69 Industrial Park Road East
Tolland ,  CT 06084-2805

APPLICATIONS

Ambient Monitor Gas Calibration with Ozone Generators - Air and Climate - Air Monitoring and Testing

Environics® Ambient Monitor Calibrator with Ozone Generators facilitate dynamic calibration of ambient air analyzers. These instruments automatically perform zero, precision, span and multi-point calibrations using NO, NO2, SO2, CO, O3, hydrocarbons and other gases of interest. All ozone generators are factory calibrated using a NIST traceable ozone standard. The Series 6100, Series 6103 and Series 9100 all meet or exceed U.S. Environmental Protection Agency requirements.