HORIBA - F-50 Series - Benchtop Multi Parameter Water Quality Meters

From Process & Environmental - Water Quality Measurement

0.511.522.533.544.55 (0 votes)
HORIBA - F-50 Series - Benchtop Multi-parameter Water Quality Meters

Years of HORIBA sensor technology development have reached their culmination in the form of a 46-mm diameter sensor probe: a compact monitoring solution offering high pressure tolerance, long-term continuous measurement capability and highly accurate, simultaneous analysis of 13 parameters.


  • F-51BW
    pH/ORP
    Standard pH meter with custom black and white LCD display.
  • F-52
    pH/ORP
    Color LCD display with navigation function.
  • F-53
    pH/ORP/ION
    Two channel meter with pH, ORP, and ISE measurment, color LCD display with navigation feature.
  • F-54BW
    pH/ORP/Conductivity
    Economical pH, ORP, and conductivity meter with custom black and white LCD display.
  • F-54
    pH/ORP/Conductivity/Salt/Resistivity
    Two channel meter with pH, ORP, conductivity, salt, and resistivity measurement.  Color LCD display with navigation feature.
  • F-55
    pH/ORP/ION/Conductivity/Salt/Resistivity
    World's first meter with FDA 21CFR Part11 support.  Measures pH, ORP, ION, conductivity, salt, and resistivity.

Features:

Features:


  • Color display and navigation function for enhanced user-friendliness.
  • Enhanced data reliability with validation feature.
  • Benchtop multi-parameter meter allows measurement of up to 4 parameters.
  • Electronic record, audit trail, and electronic signatures are security function features (FDA 21CFR Part 11 compliance).

Specification

Specification


F-51BW

  • Measuring Object & Range (Measuring Method): pH 0.00 to 14.00
    (Glass electrode method)
    (Resolution 0.01pH)
    mV (ORP) -1999 to 1999
    (Glass electrode method)
    (Resolution 1 mV)
    Temperature 0.0 to 100.0°C
    (Resolution 0.1°C)
  • Repeatability: pH ±0.01pH ±1 digit
    ORP ±1mV ±1 digit
    Temperature ±0.1°C ±1 digit
  • Power Supply: DC6V (LR6 dry cell battery)
    AC adapter 100 to 240 V 50/60Hz
  • Mass: Approx. 1 kg
  • Option:
    • RS 232C printer output
    • Battery operation possible

F-52

  • Measuring Object & Range
    (Measuring Method): pH 0.000 to 14.000
    (Glass electrode method)
    (Resolution 0.01/0.001pH)
    ORP -1999 to 1999
    (Glass electrode method) (Resolution 0.1 mV)
    Temperature 0.0 to 100.0°C (Resolution 0.1°C)
  • Repeatability: pH ±0.01pH ±1 digit
    mV (ORP) ±1mV ±1 digit
    Temperature ±0.1°C ±1 digit
  • Power Supply: AC adapter 100 to 240 V 50/60Hz
  • Mass: Approx. 1.1 kg
  • Option:
    RS 232C printer output

F-53

  • Measuring Object & Range
    (Measuring Method): pH 0.000 to 14.000
    (Glass electrode method)
    (Resolution 0.01/0.001pH)
    ORP -1999 to 1999
    (Glass electrode method) (Resolution 0.1 mV)
    ION 0.0μ to 999g/L (mol/L)
    Temperature 0.0 to 100.0°C (Resolution 0.1°C)
  • Repeatability: pH ±0.001pH ±1 digit
  • mV (ORP) ±0.1mV ±1 digit
  • Temperature ±0.1°C ±1 digit
  • ION ±0.5%F.S. ±1 digit
  • Power Supply: AC adaptor 100 to 240 V 50/60Hz
  • Mass: Approx. 1.1 kg
  • Option:
    • RS 232C printer output
    • Compact flash

D-54BW

  • Measuring Object & Range
    (Measuring Method): pH 0.000 to 14.000
    (Glass electrode method
    (Resolution 0.001pH)
    ORP -1999 to 1999
    (Glass electrode method) (Resolution 1 mV)
    Temperature 0.0 to 100.0°C (Resolution 0.1°C)
    Conductivity (AC bipolar method)
    Cell constant 100m-1 : 0.000mS/m – 19.99S/m
    Cell constant 10m-1 : 0.0μS/m – 1.999S/m
    Cell constant 1000m-1 : 0.00mS/m – 199.9S/m
    Resolution 0.05% F.S.
  • Repeatability: pH ±0.01pH ±1 digit
    mV (ORP) ±1mV ±1 digit
    Temperature ±0.1°C ±1 digit
    Conductivity ±0.5%F.S. ±1 digit
  • Power Supply: DC6V (LR6 dry cell battery)
    AC adapter 100 to 240 V 50/60 Hz
  • Mass: Approx. 1 kg
  • Options:
    • RS232C printer output
    • Battery operation possible

F-54

  • Measuring Object & Range
    (Measuring Method): pH 0.000 to 14.000
    (Glass electrode method
    (Resolution 0.001pH)
    ORP -1999 to 1999
    (Glass electrode method) (Resolution 1 mV)
    Temperature 0.0 to 100.0°C (Resolution 0.1°C)
    Conductivity (AC bipolar method)
    Cell constant 100m-1 : 0.000mS/m – 19.99S/m
    Cell constant 10m-1 : 0.0μS/m – 1.999S/m
    Cell constant 1000m-1 : 0.00mS/m – 199.9S/m
    (Resolution 0.05% F.S.)
    SALT 0.00 to 4.00 %
    (Conductivity conversion)
    (Resolution 0.01 %)
    Resistance (AC bipolar method)
    Cell constant 100m-1 : 5.00 Ωm – 199.9 kΩm
    Cell constant 10m-1 : 50.0 Ωm – 1.999 MΩm
    Cell constant 1000m-1 : 0.500 Ωm – 19.99 kΩm
    (Resolution 0.05% F.S.)
  • Repeatability: pH ±0.001pH ±1 digit
    mV (ORP) ±0.1mV ±1 digit
    Temperature ±0.1°C ±1 digit
    Conductivity ±0.5%F.S. ±1 digit
    Resistance ±0.5%F.S. ±1 digit
  • Power Supply: AC adapter 100 to 240 V 50/60 Hz
  • Mass: Approx. 1.1 kg
  • Options:
    • RS232C printer output
    • Compact flash

F-55

  • Measuring Object & Range
    (Measuring Method): pH 0.000 to 14.000
    (Glass electrode method)
    (Resolution 0.01/0.001 pH)
    ORP -1999 to 1999
    (Glass electrode method)
    (Resolution 0.1 mV)
    Temperature 0.0 to 100.0°C
    (Resolution 0.1°C)
    Conductivity (AC bipolar method)
    Cell constant 100m-1 : 0.000mS/m – 19.99S/m
    Cell constant 10m-1 : 0.0μS/m – 1.999S/m
    Cell constant 1000m-1 : 0.00mS/m – 199.9S/m
    Resolution 0.05% F.S.
    SALT 0.00 to 4.00 %
    (Conductivity conversion)
    (Resolution 0.01 %)
  • Resistance (AC bipolar method)
    Cell constant 100m-1 : 5.00 Ωm – 199.9 kΩm
    Cell constant 10m-1 : 50.0 Ωm – 1.999 MΩm
    Cell constant 1000m-1 : 0.500 Ωm – 19.99 kΩm
    (Resolution 0.05% F.S.)
  • Repeatability: pH ±0.001pH ±1 digit
    mV (ORP) ±0.1mV ±1 digit
    Temperature ±0.1°C ±1 digit
    ION ±0.5%F.S. ±1 digit
    Conductivity ±0.5%F.S. ±1 digit
    Resistance ±0.5%F.S. ±1 digit
  • Power Supply: AC adapter 100 to 240 V 50/60 Hz
  • Mass: Approx. 1.1 kg
  • Options:
    • RS232C printer output
    • Compact flash
    • Ethernet

HORIBA Europe GmbH
HORIBA Europe GmbH
Julius Kronenbergstrasse 9
Leichlingen 42799
RELATED KEYWORDS

MOST POPULAR RELATED SEARCHES

Post a new comment
Post your comment
View comments

No comments were found for HORIBA - F-50 Series - Benchtop Multi-parameter Water Quality Meters. Be the first to comment!



Air & ClimateEnergy & RenewablesEnvironmental ManagementHealth & SafetyMonitoring & TestingSoil & GroundwaterWaste & RecyclingWater & Wastewater