- T200 - Chemiluminescence NO/NO2/NOX Analyzer

From
0.511.522.533.544.55 (0 votes)
TAPI - T200 - Chemiluminescence NO/NO2/NOX Analyzer

The Model T200 NO/NO2/NOx analyzer uses the proven chemiluminescence detection principle, coupled with state-of-the-art electronics to allow accurate and dependable low level measurements for use as an ambient analyzer or dilution CEMS monitor.

A unique AutoZero feature provides superb stability by continuously correcting for zero drift, while advanced Adaptive Filtering allows the analyzer to optimize performance under changing conditions. The Model T200 includes a permeation inlet dryer for ozone generation to provide excellent reliability with no periodic replacement required. A catalytic exhaust ozone scrubber is standard for maximum safety and pump life.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports either locally or by remote connection using the included APIcom™ software.

Features

  • Ranges: 0-50 ppb to 0-20 ppm, user selectable
  • Independent ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Internal zero & span check (optional)
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty

Specification

General

  • Ranges
    • Min: 0-50 ppb Full scale
    • Max: 0-20,000 ppb Full scale (selectable, independent NO, NO2, NOx ranges and auto ranging supported)
  • Measurement Units: ppb, ppm, μg/m3, mg/m3 (selectable)
  • Zero Noise: < 0.2 ppb (RMS)
  • Span Noise: < 0.5% of reading (RMS) above 50 ppb
  • Lower Detectable Limit: 0.4 ppb
  • Zero Drift: < 0.5 ppb/24 hours
  • Span Drift: < 0.5% of full scale/24 hours
  • Lag Time: 20 seconds
  • Rise and Fall Time: < 60 seconds to 95%
  • Linearity: 1% of full scale
  • Precision: 0.5% of reading above 50 ppb
  • Sample Flow Rate: 500 cm3/min ±10%

Electrical Specifications

  • Power Requirements: 100V-120V, 220V-240V, 50/60 Hz
  • Analog Output Ranges: 10V, 5V, 1V, 0.1V (selectable)
  • Recorder Offset: ±10%

Communication Specifications

  • Included I/O
    • 1 x Ethernet: 10/100Base-T
    • 2 x RS232 (300-115,200 baud)
    • 2 x USB device ports
    • 8 x opto-isolated digital outputs
    • 6 x opto-isolated digital inputs
    • 4 x analog outputs
  • Optional I/O
    • 1 x USB com port
    • 1 x RS485
    • 8 x analog inputs (0-10V, 12-bit)
    • 4 x digital alarm outputs
    • Multidrop RS232
    • 3 x 4-20mA current outputs

Physical Specifications

  • Operating Temperature Range: 5 - 40ºC (with EPA Equivalency)
  • Dimensions (HxWxD): 7” x 17” x 23.5” (178 x 432 x 597 mm)
  • Weight
    • Analyzer: 40 lbs (18 kg)
    • External pump: 15 lbs (7 kg)

Certifications

  • US EPA: RFNA-1194-099
  • MCERTS: Sira MC 050068/05
Teledyne Advanced Pollution Instrumentation, Inc. (TAPI)
Teledyne Advanced Pollution Instrumentation, Inc. (TAPI)
9480 Carroll Park Drive
San Diego ,  California 92121-5201