- Model T400 - UV Absorption O3 Analyzer

From
0.511.522.533.544.55 (0 votes)
TAPI - Model T400 - UV Absorption O3 Analyzer

The Model T400 UV Absorption analyzer uses a system based on the Beer-Lambert law for measuring low ranges of ozone in ambient air.

A 254 nm UV light signal is passed through the sample cell where it is absorbed in proportion to the amount of ozone present. Periodically, a switching valve alternates measurement between the sample stream and a sample that has been scrubbed of ozone. The result is a true, stable ozone measurement.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports either locally or by remote connection using the included APIcom™ software.

Features

  • Ranges: 0-100 ppb to 0-10 ppm, user selectable
  • Single pass ultraviolet absorption
  • Dual ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Internal programmable zero, span and precision points (optional)
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty

Specification

General

  • Ranges
  • Min: 0-100 ppb Full scale
  • Max: 0-10 ppm Full scale (selectable, dual ranges and auto ranging supported)
  • Measurement Units: ppb, ppm, μg/m3, mg/m3 (selectable)
  • Zero Noise: < 0.3 ppb (RMS)
  • Span Noise: < 0.5% of reading (RMS) above 100 ppb
  • Lower Detectable Limit: < 0.6 ppb
  • Zero Drift: < 1.0 ppb/24 hours
  • Span Drift: < 1% of reading/24 hours
  • Lag Time: < 10 seconds
  • Rise and Fall Time: < 20 seconds to 95%
  • Linearity: 1% of full scale
  • Precision: <0.5% of reading above 100 ppb
  • Sample Flow Rate: 800 cm3/min ±10%
  • IZS Specifications
  • Maximum Concentration: 1.0 ppm
  • Minimum Concentration: 0.050 ppm
  • Resolution: 0.5 ppb
  • Repeatability (7 days): 1% of reading
  • Initial accuracy: ± 5% of target

Electrical Specifications

  • Power Requirements: 100V-120V, 220V-240V, 50/60 Hz
  • Analog Output Ranges: 10V, 5V, 1V, 0.1V (selectable)
  • Recorder Offset: ±10%

Communication Specifications

  • Included I/O
    • 1 x Ethernet: 10/100Base-T
    • 2 x RS232 (300-115,200 baud)
    • 2 x USB device ports
    • 8 x opto-isolated digital outputs
    • 6 x opto-isolated digital inputs
    • 4 x analog outputs
  • Optional I/O
    • 1 x USB com port
    • 1 x RS485
    • 8 x analog inputs (0-10V, 12-bit)
    • 4 x digital alarm outputs
    • Multidrop RS232
    • 3 x 4-20mA current outputs

Physical Specifications

  • Operating Temperature Range: 5 - 40ºC (with EPA Equivalency)
  • Dimensions (HxWxD): 7” x 17” x 23.5” (178 x 432 x 597 mm)
  • Weight
    • 28 lbs (12.7 kg)
    • 30.6 lbs (13.8 kg) with IZS

Certifications

  • US EPA: EQOA-0992-087
  • MCERTS: Sira MC 050070/04
Teledyne Advanced Pollution Instrumentation, Inc. (TAPI)
Teledyne Advanced Pollution Instrumentation, Inc. (TAPI)
9480 Carroll Park Drive
San Diego ,  California 92121-5201

DOWNLOADS RELATED TO THIS PRODUCT

MOST POPULAR RELATED SEARCHES

Post a New Comment

0 Comments

No comments were found for TAPI - Model T400 - UV Absorption O3 Analyzer. Be the first to comment!

Environmental XPRT is part of XPRT Media All Rights Reserved