Contact Now/Request quote View Phone/Fax Number Visit Website

HORIBA - OCMA-310 - Oil Content Monitor

0.511.522.533.544.55 (0 votes)
HORIBA - OCMA-310 - Oil Content Monitor

The OCMA-310 is an automated portable unit for monitoring oil content in both fresh water and saltwater. The one-touch monitoring process is fully automated, from sample extraction to analysis and sample expulsion. The OCMA-310 utilizes an S-316 extraction solvent to ensure superior oil particle stability and infrared absorptiometry for sample analysis. Using 0mg/l -200mg/l dynamic ranging, the unit is capable of analyzing both high and low concentrations. The OCMA-310 provides swift, highly accurate analysis of the oil content of both fresh water and saltwater as well as that of such fluids as those found at factories and wastewater treatment plants.

Features

Features


  • A microcomputer provides automated analysis of the samples at the touch of a button.
  • Dynamic ranging provides analysis from 0mg/l-200mg/l without manual switching.
  • The use of an S-316 extraction solvent ensures the highest levels of oil particle stability.
  • A digital display eliminates the misreading of analysis results. A visual display alerts users to the unit's operating status.
  • The use of a proprietary syringe ensures a fixed amount of sample water and solvent.


Stable readout and a broad dynamic range assure accuracy and simplify analysis of oil components.

The new OCMA-310 Oil Content Analyzer can assess the amount of trace hydrocarbons in soil or water, as well as check for residual hydrocarbons on pre-cleaned products. The extraction of the oil can be done by Freon 113 and S-316. The S-316 solvent used by the unit may be recycled by HORIBA’s optional solvent reclaimer, making the OCMA-310 very environmentally-friendly. Furthermore, the unit’s broad measurement range eliminates the need for troublesome range switching. As a result, the OCMA-310 is quick, easy to use, accurate, and reliable.

Manual control
The two valve system lets you make sure that your sample is measured properly. Inject your sample and press the extract button to initiate mixing and separation. You can visually confirm mixing and separation through the sample window. Open the upper valve (extraction valve), and the sample will flow into the measuring cell. An oil/water separation filter prevents water from entering the cell. An overflow system discharges any excess sample. The lower valve (drainage valve) allows you to drain the sample after it has been measured.

0-200 mg/l dynamic range
The OCMA-310’s pyro-electric sensor gives you a full dynamic range from 0-200 mg/L. This eliminates the need for range-switching and provides fast, accurate measurements for all types of samples, regardless of concentration levels.

Quick start-up
The OCMA-310 warms up in less than 30 minutes, enabling you to get started quickly.

No optical adjustments
The troublesome optical adjustments commonly required by conventional pneumatic NDIR analyzers are no longer necessary.

Auto hold function
The unit’s internal computer automatically determines when the measured value has stabilized and maintains that value on the display.

Self-diagnosis
The OCMA-310 monitors itself for problems such as motor malfunction, deterioration of the measuring cell’s light source, and electrical trouble. Should anything out of the ordinary occur, an alarm light will turn on, and an appropriate error message will appear on the operation display panel.

Oil/water separation filter
The OCMA-310 is equipped with an oil/water separation filter which prevents water and dirt from entering the measuring cell. This prevents the cell from being damaged, while insuring that the sample being measured is not contaminated.


Recommended Applications

Recommended Applications


  • Environment: Surveying environmental water quality and hazardous waste sites
  • Industry: Monitoring of waste-water influent and effluent
  • Marine transportation: Checking bilge and ballast discharge
  • Petroleum processing: Monitoring the efficiency of oil/water separation processes
  • Oil depots: Monitoring the discharge produced when cleaning storage tanks
  • Automotive: Monitoring waste-water discharge from service stations
  • Quality Assurance: Measuring residual oil on textiles, metal parts, etc.

Specification

Specification


  • Measurement principle: Solvent extraction, NDIR analysis (Infrared spectroscopy)
  • Measurement range: 0mg/l ~ 200mg/l dynamic ranging
  • Resolution: 0 to 99.9 mg/L; 0.1 mg/L
    100 to 200 mg/L; 1 mg/L
  • Repeatability: 0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit
    10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
    100 to 200 mg/L; ± 4 mg/L ± 1 digit
  • Measurement: Manually controlled
  • Calibration: One touch calibration after the calibration standard is introduced to the instrument
  • Extraction Solvent: S-316 solvent
  • Sample/Solvent Volume: 20 mL, ratio of sample to solvent is 2:1.
  • Extraction Method: Built-in extractor for oily water samples
  • Display: Measured value; 3 digits LCD with backlight Message; Character display LCD with back-light (16×2 characters)
  • Functions: Data memory up to 50 data, Self diagnostics, Stable type display, Interactive operation, Calendar clock
  • Output: RS-232C, parallel printer port
  • Option: Analog output 0 to 200 mg/L;0 to 1 V
  • Ambient Temperature: 0 to 40 ℃
  • Power: 100 to 120 V AC ±10%,50/60 Hz, 80 VA
    200 to 240 V AC ±10%,50/60 Hz, 120 VA
  • Dimensions: 200(W)×315(D)×342(H) mm
  • Mass: Approx. 7 kg

HORIBA Europe GmbH
HORIBA Europe GmbH
Julius Kronenbergstrasse 9
Leichlingen 42799
DOWNLOADS

DOWNLOADS


RELATED KEYWORDS

MOST POPULAR RELATED SEARCHES

Post a new comment

Post your comment

View comments

No comments were found for HORIBA - OCMA-310 - Oil Content Monitor. Be the first to comment!



Contact Us | Advertise | FAQs | Add Your Company | About Us | Terms | Privacy Policy | Company Sitemap | Category Sitemap | Professional Profiles Sitemap

Environmental Expert LinkedIn LinkedIn | Environmental Expert Twitter Twitter | Environmental Expert Facebook Facebook | RSS Feeds RSS | Environmental Expert Blog Blog

©1999-2012 Environmental Expert S.L.
Air & ClimateEnergy & RenewablesEnvironmental ManagementHealth & SafetyMonitoring & TestingSoil & GroundwaterWaste & RecyclingWater & Wastewater