Each method involves the measurement of a parameter related to particulate concentration and then requires a calibration to be determined to convert the measured value into one of particulate mass concentration.
In order to output particulate concentration in mg/m3, a Density Scale Factor (DSF) must be determined. The Density Scale Factor for any given installation will vary based on the type, shape, size and colour of the particles in the gas stream and therefore, the Density Scale Factor must be determined on site, through comparison of the instrument readings with an independent reference measurement*.
During normal operation:
Particulate concentration (mg/m3) = measured parameter x DSF
*It is recommended that standard reference measurement is made using gravimetric analysis based on isokinetic sampling of the gas stream such as EN13284-1:2002 or US EPA Method 5.