Procedure for extended range atrazine RaPID Assay, installation on the RPA-I

0

Courtesy of

General Comments

New RaPID Assay methods can be entered into the RPA-I memory with either a download cartridge or manually. This Technical Bulletin is a guide to the manual installation of extended range atrazine method on the RPA-I using the keypad. Please refer to the RPA-I Manual for a detailed discussion.

Protocol

1. Turn on the RPA-I.
2. Press the [OPTION] key on the top right side of the keypad.
3. Scroll through the menu by pressing either the [YES] or [NO] keys. Press the [ENTER] key when 'Enter/Edit Proto' appears.
4. Scroll through the menu using the [YES] or [NO] keys until 'UNUSED [# Avail.]' appears. The '# Avail'indicates how many more protocols the cartridge will accept. Press the [ENTER] key (if 'NO' appears, a protocol must be deleted before a new one can be entered).
5. Protocol Nam: Enter the name of the new protocol 'XTND-ATZ' by scrolling with either the [YES] or [NO] keys until letter 'X' appears. Press the [ENTER] key. Scroll to 'T', press [ENTER]. Continue inputting letters until the entire name is entered. Errors can be corrected by pressing the [BACK SPACE] key.
6. Data Reduct.: Select by scrolling to Lin. Regression and press [ENTER].
7. Xform Option: Select Ln/LgtB and press [ENTER].
8. # of Calibrator: Enter 4 and press [ENTER].
9. Cal. Replicates: Enter 2 and press [ENTER].
10. Cal. #1 Conc: Enter 0.00 and press [ENTER].
11. Cal. #2 Conc: Enter 0.5 and press [ENTER].
12. Cal. #3 Conc: Enter 2.5 and press [ENTER].
13. Cal. #4 Conc: Enter 10.0 and press [ENTER].
14. Correlation Flag: Enter 0.990 and press [ENTER].
15. Unit: Enter 'PPB' as in step 5 and press [ENTER] after each letter. Press [ENTER] a second time after the second 'B' to continue.
16. Normal Range Hi: Enter 10.0 and press [ENTER].
17. Normal Range Low: Enter 0.1 and press [ENTER].
18. # of Controls: Enter 1 and press [ENTER].
19. Ctrl Replicates: Enter 1 and press [ENTER].
20. # of Rgt Blanks: Enter 0 and press [ENTER].
21. Wavelength: Enter 450 and press [ENTER].
22. Read Mode: Scroll to Absorbance and press [ENTER].
23. Rep. %CV Flag: Enter 10 and press [ENTER].
24. List Protocol?: Press the YES key. Check the data entered with the information listed here.