`An Inspector Calls`

0

Source: Pollution & Process Monitoring Ltd

As part of an order for nine instruments placed by our Indian distributer, three Protoc analysers were independently inspected by Lloyds Register as a proportional sample of the batch order. The inspection followed a Quality Assurance Plan request by the end user installing the instruments at a refinery project, measuring condensate and cooling water.

The three Protoc TOC instruments were ranged 0-10, 0-20 and 0-50 PPM TOC and were verified for linearity, 24 hour stability and the analogue and digital outputs were also validated. The remaining six instruments were performance evaluated by PPM engineers prior to the visit so the test paperwork could be acknowledged by the Lloyds inspector.