airmoCAL - Calibration Device
From Peripherals - Calibration
Calibration device inside 19? and 4U rack. Designed for automatic validation of results at different concentration levels. ZERO / Sample / Cylinder / CALIB.
Principle
The airmoCAL is designed for automatic validation of results. The calibration is ensured by permeation tubes (e.g. butane, hexane, benzene…) and/or by cylinder.
The permeation tubes swept by a constant pure air stream are installed in a precisely regulated temperature oven. These two parameters allow a full control of the emitted gas concentration.
A certified cylinder can also be used for the calibration and validation of results.
The system is controlled by Vistachrom software and 4 methods can be programmed for the calibration: ZERO, Cylinder, sample and CALIB.
Multiplexer Option:
This multiplexer is made with inert tubes and will be linked to different lines (one line with one METHOD,). The total number of methods is 16 (multiplexer + CALIB + Cylinder + … + …)
It allows to analyse on sequence or to choose any of the different METHOD in manual mode.
It purges the different lines that are not in analysis to lower the sampling time for the next analysis and to avoid for any compound to be adsorbed in the line.
To check the stability of the flow, a flow meter is installed on the front panel. It is important to control the flow every year with a flowmeter calibrated at +/-1%.
VISTACHROM software:
Chromatotec® developed software system which enables:
- Remote control
- Full traceability with on board archiving of results
- Programation and control of threshold alarms
- Ponctual (laboratory) or continuous measurement
- Data export by MODBUS / JBUS / MGS1 / 4-20mA / 0-10V
Options
- airmoCAL-D for several dilution points
- Gas purifier
- Module 4 x analog output 4-20 mA or 0-10 V.
- Alarms module
- Automatic results validation.
- 24 V power supply.
- Hydrogen and Nitrogen generator
- Sampling pump in 24V DC or airmoPUMP (XXX901 or XXX915)
- Pressure and temperature measure
- Mass Flow Controller (MFC, in order to improve dilution precision and so calibration quality)
Technical Specifications:
Permeation tubes available:
- Butane, hexane, benzene in standard for VOC applications
- Others upon request
Oven temperature:
- 45°C ± 0.1°C in standard
Air stream flow:
- Depending on the expected calibration concentration
Programmation:
- The airmoCAL is fully controlled by Vistachrom software
Cycle time :
- The same than analysis
Gas supply:
- Air
Power supply:
- Main (230V / 115V 50 Hz/60Hz)
- Battery 24V (option)
Electrical consumption:
- Average : 100 VA
- Peak : 200 VA
Dimensions and weight:
- Rack: (19ˮ) 4U
- Height: 177 mm
- Width: 482 mm
- Depth: 660 mm
- Net Weight: 20 kg
- If option airmoCAL-M: rack 5U