Aquaprobe - Model AP-5000 - Advanced Mulitparameter Mini Water Quality Sonde Aquaprobe
From Water Quality
The AP-5000 is the larger of our advanced portable multiparameter water monitoring probes. It comes with a range of standard sensors included, and much like its smaller counterpart it offers more customisation options that allow you to add extra sensors to the probe. All of our probes are constructed from tough marine grade aluminium that has been hard anodised, making it suitable for use in both fresh and salt waters.
-
Most popular related searches
Standard Parameters on the AP-5000
- Optical DO
- Specific EC
- Absolute EC
- pH
- ORP
- TDS
- Resistivity
- Salinity
- SSG
- Temperature
- Depth
The AP-5000 features 4 Auxiliary sockets to allow you to add further sensors to the probe. Unlike the AP-2000, these auxiliary ports are unrestricted. This means that each port can take either an Ion Selective Electrode (ISE) or an optical sensor. This makes the number of potential sensor combinations vast.
The probe can be used with a variety of cable lengths via the AquaConn Connectors. Cables available in 3, 10, 20 and 30m lengths as standard.
Customisation Options
- Customising the AP-5000 is simple, there are 4 ports and you can add any combination of our ISE and optical sensors into these ports.
- Probes are supplied with blanking plugs in the auxiliary ports as standard. Simply unscrew these to add your extra electrodes.
Using the AP-5000 as a Water Level Logger
The AP-5000 features a depth sensor. This sensor is capable of measurement accuracy of 0.5%FS. This is very comparable to a dedicated water level logger such as the LeveLine-ECO: 0.2%FS, and the LeveLine (titanium): 0.1%FS.
The depth is zeroed automatically whilst the conductivity is reading zero. As soon as the conductivity registers over 0 the probe depth begins to be calculated. Once the probe is lowered to the desired depth it must be secured, then using the GPS Aquameter the depth can be zeroed once again. Now the AP-5000 will detect small changes in the water level and display them as a positive or negative change.
Some of the AP-5000’s Features
Advanced Optical DO sensor
The optical sensor works on the principle of Dynamic Luminescence Quenching. A gas-permeable material known as a luminophore is excited with short bursts of blue light, which causes molecules in the luminophore to emit red photons. By measuring the delay of the returned red photons with respect to the blue excitation, the level of dissolved oxygen present can be determined.
The optical method has various advantages over the historical galvanic method for measuring dissolved oxygen.
The most important being that as no oxygen is consumed across a membrane, the sensor does not require a flow of liquid passing over it to achieve a stable reading. Other advantages include in frequent calibrations (every 3-6 months), and replaceable caps that last over 2 years.
Load More than One Optical Sensor
Load More than One Optical Sensor
The AP-5000 has four auxiliary sockets. These sockets allow you to add any combination of extra sensors to the AP-5000 depending on your requirements.The main advantage the AP-5000 has over the AP-2000 is that you can install 2 optical sensors such as turbidity and chlorophyll for example. We have a wide range of sensors available including ammonium, nitrate, blue-green algae and refined oil to name a few.
Rugged Connection
Rugged Connection
All Aquaprobes feature the new AquaConn fully metal connectors. The AP-5000 features the all-metal AquaConn connectors. These high specification connectors feature large metallic key for a rigid fit, recessed pins for extra protection during pairing, guide markers on each part of the connector to allow you to instantly push the pair together in the correct alignment.
Technical Specifications
Dissolved Oxygen:
- Range: 0 – 500.0% / 0 – 50.00 mg/L
- Resolution: 0.1% / 0.01mg/L
- Accuracy: 0 – 200%: ± 1% of reading. 200% – 500%: ± 10%
Conductivity (EC):
- Range: 0 – 200 mS/cm (0 – 200,000 µS/cm)
- Resolution: 3 Auto-range scales: 0 – 9999 µS/cm, 10.00 – 99.99 mS/cm, 100.0 – 200.0mS/cm
- Accuracy: ± 1% of reading or ± 1µS/cm if greater
TDS*:
- Range: 0 – 100,000 mg/L (ppm)
- Resolution: 2 Auto-range scales: 0 – 9999mg/L, 10.00 – 100.00g/L
- Accuracy: ± 1% of reading or ± 1mg/L if greater
Resistivity*:
- Range: 5 Ωcm – 1 MΩcm
- Resolution: 2 Auto-range scales: 5 – 9999 Ωcm, 10.0 – 1000.0 KΩcm
- Accuracy: ± 1% of reading or ± 1 Ωcm if greater
Salinity*:
- Range: 0 – 70 PSU / 0 – 70.00 ppt (g/Kg)
- Resolution: 0.01 PSU / 0.01 ppt
- Accuracy: ± 1% of reading or ± 0.1 unit if greater
Seawater Specific Gravity*:
- Range: 0 – 50 st
- Resolution: 0.1 st
- Accuracy: ± 1.0 st
pH:
- Range: 0 – 14 pH / ± 625mV
- Resolution: 0.01 pH / ± 0.1mV
- Accuracy: ± 0.01 pH / ± 5mV
ORP:
- Range: ± 2000mV
- Resolution: 0.1mV
- Accuracy: ± 5mV
Temperature:
- Range: -5°C – +50°C (23°F – 122°F)
- Resolution: 0.01° C/F
- Accuracy: ± 0.5° C
Optional Parameters
Depth **
- Range: 0-60m (AP-2000 & AP-5000), 0 – 100m (AP-7000)
- Resolution: 0.01m
- Accuracy: ±0.5% FS (AP-2000 & AP-5000), ±0.2%FS (AP-7000)
**Option on the AP-2000. Comes as standard on the AP-5000 & AP-7000Ion Specific Electrodes***
Ammonium:
- Range: 0 – 9,000mg/L (ppm)
- Resolution: 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L
- Accuracy: ± 10% of reading
Ammonia:
- Range: 0 – 9,000mg/L (ppm)
- Resolution: 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L
- Accuracy: ± 10% of reading
Chloride:
- Range: 0 – 20,000mg/L (ppm)
- Resolution: 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 20,000 mg/L
- Accuracy: ± 10% of reading
Flouride:
- Range: 0 – 1,000mg/L (ppm)
- Resolution: 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
- Accuracy: ± 10% of reading
Nitrate:
- Range: 0 – 30,000mg/L (ppm)
- Resolution: 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 30,000 mg/L
- Accuracy: ± 10% of reading
Calcium:
- Range: 0 – 2,000mg/L (ppm)
- Resolution: 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 2,000 mg/L
- Accuracy: ± 10% of reading
Ammonium electrode required. Readings calculated from Ammonium, pH and temperature electrode values.
Optical Electrodes***
Turbidity:
- Range: 0 – 3000 NTU
- Resolution: 2 Auto-range scales: 0.0 99.9 NTU, 100 – 3000 NTU
- Repeatability: ± 2% of reading
Chlorophyll:
- Range: 0 – 500 µg/L (ppb)
- Resolution: 0.01 µg/L
- Repeatability: ± 2% of reading
Phycocyanin (Freshwater Blue-Green Algae):
- Range: 0 – 300,000 cells/mL
- Resolution: 1 cell/mL
- Repeatability: ± 2% of reading
Phycoerythrin (Marine Blue-Green Algae):
- Range: 0 – 200,000 cells/mL
- Resolution: 1 cell/mL
- Repeatability: ± 2% of reading
Rhodamine WT Dye:
- Range: 0 – 500 µg/L (ppb)
- Resolution: 0.01 µg/L
- Repeatability: ± 5% of reading
Fluorescein Dye:
- Range: 0 – 500 µg/L (ppb)
- Resolution: 0.01 µg/L
- Repeatability: ± 5% of reading
Refined Oil:
- Range: 0 – 10,000 µg/L (ppb)
- Resolution: 0.1 µg/L
- Repeatability: ± 2% of reading
CDOM/FDOM:
- Range: 0 – 2,000mg/L (ppm)
- Resolution: 2 Auto-range scales:
- 0.00 – 99.99 mg/L, 100.0 – 2,000mg/L
- Repeatability: ± 10% of reading or 2ppm (whichever is greater)
***Note
In addition to the standard parameters supplied with the base Aquaprobe, the AP-2000 has 1 ISE port and 1 optical port, the AP-5000 has 4 ports that can take any combination of ISE or optical sensors and the AP-7000 has 6 ports that can also take any combination of ISE or optical sensors.
The accuracy figures quoted above represent the equipment’s capability at the calibration points at 25°C. These figures do not take into account errors introduced by variations in the accuracy of calibration solutions and errors beyond the control of the manufacturer that may be introduced by environmental conditions in the field.
Customer reviews
No reviews were found for Aquaprobe - Model AP-5000 - Advanced Mulitparameter Mini Water Quality Sonde Aquaprobe. Be the first to review!