F&J SPECIALTY PRODUCTS, INC.

- Model DF-60810-MHVE - Mega High Volume Air Sampler System

From

0

The DF-60810-MHVE Series Air Sampling Systems are designed for remote unattended continuous air sampling applications. The DF-60810-MHVE Series Air Samplers feature a brushless motor with electronic motor speed control that maintains a user selectable flow rate. The flow rate attainable through the filter media is dependent upon the air porosity of the filter media. Flow rate as high as 200 CFM (340 m3/hr) are attainable with glass fiber filter media. The DF-60810-MHVE Series design accommodates rapid field service and component replacement. The basic components of the system are assembled in a modular fashion so that each component can be readily and independently removed for service.

For durability and weather resistance, the system is housed in a freestanding powder coat painted aluminum enclosure. The sample air is drawn in under the eaves of the hinged lid from all four sides and is exhausted near the bottom of the enclosure. The locking swing door on the enclosure provides convenient access for servicing the equipment inside. A lockable latch on the top cover restricts unauthorized tampering with the filter holder.

The electronic flow control measurement sub-system of the DF-60810-MHVE Series provides a reference standard flow measurement and an operator selectable constant flow of air through the filter medium. The air flow is measured by a precision-machined differential pressure sensor. The controller can be readily set to any sampling flow rate between 50 and 200 CFM (84-340 m3 /hr) depending on the filter paper air resistance and dimensions. The bright LED readout displays multiple air sampling information including current flow rate, current elapsed sample time and totalized volume. The filter holder can be custom designed to accommodate any filter size and type. The DF-60810-MHVE standard model utilizes an 8”×10” (20,3×25,4 cm) filter.

  • Display in English or metric units set at factory
  • Choices of flow/volume units:
    • SLPM SL
    • SCMH SCM
    • SCFM SCF
  • State of the Art microprocessor electronics
  • Automatic Flow Control
  • Auto Shut-off on time or volume
  • Flowrate and volume totalizations displayed are corrected to a factory settable Reference Temperature and Pressure (4 options available)
  • Elapsed time meter
  • Auto zero calibration feature of flow sensor
  • Bright LED display
  • Flow rate accuracy within ± 4.0% F.S.
  • RS-232 Communication Port w/Operator selectable download frequency for real-time data collection or transmission
  • 200 – 240 VAC, 50/60Hz; single phase

  • Performance: Basic components of the system are modular and independently serviceable. Sample flow rate can be set between 5 and 200 CFM (84 and 340 m3 /hr). The standard filter holder has the dimensions 20,3×25,4 cm (8”×10”).
  • Technology: Microprocessor controlled state of the art electronics
  • Operating Temperature Range: -31ºF* to 122ºF (-35ºC* to 50ºC)  *warm start/continuous operation
  • Operating Relative Humidity: 0 – 95% RH
  • Typical Flow Rate Range: 50 – 200 CFM (84 to 340 m3/hr) (Depending on filter paper dimensions and air resistance).
  • Motor: Brushless: 2.4H.P.(1800 Watt) motor with electronic motor speed control
  • Power Requirements: 220-240VAC; 50/60Hz; 16 amperes; single phase.
  • Housing: Powder coat painted aluminum Locking hinged cover
  • Removable hinged cover Locking swing door with key
  • Dimensions: 57.5”H × 21.5”W × 21.5”D (146 H × 54,6 W × 54,6 cm D)
  • Weight: Approximately 98 lbs. (44,5 kg)
  • Shipping Weight: Approximately 150 lbs. (68,2 kg)
  • Installation Category: Pollution Degree 3
  • Enclosure Rating: IPX3
  • Automatic Flow Control: The system microprocessor monitors flow rate relative to the preset Reference T and P flow rate established during the setup procedure and electronically adjusts the electronic motor speed adjustment, if necessary, to maintain the flow within ± 4% of setting. The microprocessor computes the STP flow rate by correcting the measured values of the Reference values. 

  • Measurements:
    • Temperature of air flow through system
    • Inlet pressure to the flow sensor
    • Differential Pressure of the flow sensor
  • Calculations/Determinations:
    • Totalized volume, Reference T and P
    • Current flow rate, Reference T and P
    • Elapsed time
  • Factory Settable Reference T and P
    • Classical STP 0ºC, 101.325 kPa
    • Normal T and P 20ºC, 101.325 kPa
    • Modified Normal T and P 70ºF, 101.325 kPa
    • Standard Ambient T and P 25ºC, 101.325 kPa
  • Other System Features:
    • Automatic shut off of system on totalized volume or elapsed time
    • RS-232 port for real-time data download
    • Utilization of 8”×10” (20,3×25,4 cm) filters
    • Bright LED display
    • Automatic flow control
  • Data Storage System:
    • Data Storage Device (P/N: 232FCDSD)
    • 2 GB Secure Digital Card (P/N: 372239)
    • Options: Flash Card Reader (P/N:SDDR-199-A20)

Customer reviews