F&J SPECIALTY PRODUCTS, INC.

- Model GAS-EDL-300WE-HT - Elite Digital Light (EDL) Global Air Sampling System (200 - 240 VAC)

From

F&J Model GAS-EDL-300WE-HT air sampler is a powerful small foot print light weight AC/DC powered air sampling system operable from (1) line power or (2) an optional 24VDC external power source. This powerful air sampler is the lightest weight low volume air sampler available. The HT model is designed for operation in ambient temperatures as high as 60°C (140°F).

The GAS-EDL-300WE-HT air sampler is well suited for various types of air sampling application such as REMP, plant emissions and work area monitoring programs.  It can also be utilized in remote air sampling applications powered from solar panels or wind turbines. The Global Air Sampler software provides the ultimate in end user customization including operator selectable alarm settings, Reference T and P values and engineering units.

The F&J technology provides maximum flexibility, data integrity, automatic flow control and opportunities for control by a PC device through the RS232 communication port.

The typical operating flow range is 30 – 250 SLPM (1.1 – 8.8 SCFM). Actual flow range implemented will depend the filter(s) utilized by the customer.

  • State-of-the-Art Electronics
  • Lightweight: ~15 lbs. (6,8 kg)
  • Powerful: Flow rate to 300 LPM*
  • Very low maintenance
  • Automatic Flow Control
  • Precision machined DP flow sensor
  • Vacuum Fluorescent Display (VFD); 4 line × 24 characters
  • Dual RS-232 communication ports
  • Flow rate accuracy: ±3.0% Full Scale
  • Auto zero calibration feature of flow sensor
  • Multiple operator selectable data storage and data transmission frequency rates
  • Wide temperature range electronics
  • Brushless motor
  • High ambient temperature rating to 140°F (60°C)

  • Pump Type: Centrifugal blower
  • Maximum Flowrate:
    • 225 LPM - Typical w/47mm FP47M glass fiber media
    • 300+ LPM – Free air flow capacity
  • Power Source:
    • Line Power; 200VAC to 240VAC
    • External DC Source – 24VDC
  • Current Draw: 14 A maximum
  • Filter Holder Fitting:
  • 3/8 FNPT quick disconnect
  • Handle: Durable plastic
  • Weight: 18 lbs. (8.2 kg)
  • Dimensions: 10” × 11” × 14” (26 × 28 × 36 cm)
  • Operating Temperature Range:
  • 14°F to 140°F (-10°C to 60°C)
  • Display: VFD, 4×24 characters
  • Flow Control:
  • Adjustable from keypad between 30-250 LPM
  • Flow Accuracy: ±3.0% of Full Scale
  • Communication Interface: Dual RS-232

  • Language Options: English
  • Sampling Mode: Volumetric Flow or Mass Flow
  • Gas Type: Air, O2,N2,H2, CO, CO2, C3H6, He, NH3
  • Engineering Units
    • Volumetric Flow: sccm, SLPM, SCFM, sm3/min, sm3/hr
    • Mass Flow: kg/hr, g/min, 1 lbs/hr
    • Temperature: °C, °F
    • Pressure: In, Hg, mm Hg, bar, atm, kPa, hPa
  • Reference T and P
    • Reference T: 0°C, 15°C, 20°C, 21.1°C (70°F), 25°C
    • Reference P: 101.325 kPa (760 mm Hg), 100 kPa (1 bar)
  • RS232 Data Output Frequency: 1 sec, 1 min, 10 min, 20 min, 30 min, 1 hr
  • Data Storage Frequency: 1 min, 10 min, 20 min, 30 min, 1 hr
  • Operating Mode: Continuous, Periodic
  • Periodic Sampling Options: 1 hr. (12 five minute periods), or weekly (24 one hour periods for 7 days)
  • Ending Mode: By time, By volume
  • Operator Selectable Passwords: 2 levels
  • Date and Time Setup: Input of real time and date

  • Air Sampler Data Acquisition Program P/N: GASdaq
  • Isokinetic sampling mode configuration P/N: ISOKOPT
  • Ruggedized Cellular Phone System (P/N: CASRPS)