GasMix - Model AIOLOS II - Dynamic Gas Mixer
To save money and gain in flexibility, many laboratories use Aiolos – dynamic gas mixer from AlyTech – to perform multi-point calibration,linearity checks, LOQ/LOD validation. By mixing and / or diluting two to four gases the system generates calibration gas on site at the desired concentration, even to very low levels. The professional software, accuracy and reproducibility of the prepared gas standards make it an accomplished tool, well-accepted by users concerned with metrological performances.
-
Most popular related searches
Products Details
Using GasMix ™ Aiolos II you benefit of :
- Dedicated, user-friendly and intuitive software
- Less time consuming: programmed injection sequences is a great time-saver for routine quality procedures
- Full automation: by synchronizing dilutions, injection, flushing timings with an external analyzer
- Improved accuracy: multipoint MFC calibration curves for each used gas
- Rigorous metrology : full traceability, documentation, certification of the concentration accuracy
- Compliance with ISO 6145-7
Application
Who Benefits from using the AIOLOS II ?
- Environmental – Calibration of analysers for ambient air monitoring
- Chemical, Oil & Gas – Calibration of GC, μGC, Elemental analyzers, RGA, NGA; LPG analyzers, Spectrometers
- Research: Generation complex gas mixtures
- Specialty Gases – Customized secondary standards
- Food & Flavour – Quality control
Technical Specifications
Number of gas channels: 2 to 4
- Inlet Gas Pressure: 3 bars nominal, max 8 bars
- Outlet Pressure: atmospheric, higher on request
- Connections: Swagelok 1/8′ ,other on request
- Gas compatibility: pure gases, concentrated gas mixtures, most of aggressive and reactive gases (for specifics, contact us)
Performance characteristics
- Mixing : mixing up to 4 channels simultaneously
- Dilution ratio of the initial concentration: up to 1:10000
- MFC operation range: 2 to 100% FS
- Flow accuracy: ±1% of setpoint all over the scale
- Repeatability: less than 0.1% of setpoint
- Response time to the setpoint
Other
- Operation temperature: 10 – 50°C
- Power supply: 90-260Vac, 1,5-4A, 50-60Hz
- Dimensions : L x l x h (cm) : 43x25x46;
- Weight: approx. 14 kg (depends on configuration)
Customer reviews