- Model APHA-370 - Ambient THC Monitor / Ambient Air Monitor

From

0

The APHA-370 continuously monitors atmospheric THC, NMHC, and CH4 concentrations using a cross-flow modulated selective combustion type method combined with a hydrogen ion detection method. The APHA-370 employs an independent, internal dry-method sampling device to achieve the highest levels of sensitivity and accuracy.

The APHA-370 uses a combination of the flame ionization detection method and selective-combustion. This gives it the advantage of the single-detector method plus the ability to perform continuous measurements, free of zero-drift, for THC, NMHC, and CH4. The design gives great stability and high sensitivity (0-5 ppm F.S.)

The APHA-370 has a relative-sensitivity correction function for CH4 and NMHC.

All the necessary features are built right into a single rack-sized instrument, including a catalytic unit for selective combustion (i.e., an NMHC cutter); a cataytic unit for generating reference gas and auxiliary combustion air (standard); and a sampling pump. The only supplemental gas required is H2.

  • The APHA-370 uses a combination of the flame ionization detection method and selective-combustion.
  • This gives it the advantage of the single-detector method plus the ability to perform continuous measurements, free of zero-drift, for THC, NMHC, and CH4. The design gives great stability and high sensitivity (0-5 ppm F.S.)
  • The APHA-370 has a relative-sensitivity correction function for CH4 and NMHC.
  • All the necessary features are built right into a single rack-sized instrument, including a catalytic unit for selective combustion (i.e., an NMHC cutter); a cataytic unit for generating reference gas and auxiliary combustion air (standard); and a sampling pump. The only supplemental gas required is H2.

Flame ionization detection method (FID) with selective-combustion

The flame Ionization detection method (FID) - used in combination with the selective-combustion system utilizes the ionization that occurs as the result of the high-temperature energy from combustion at the tip of the burner jet when organic carbon compounds are introduced into the hydrogen flame. The hydrogen flame is located between two electrodes.

When an electrical voltage is applied across these electrodes a minute ion current proportional to the hydrocarbon concentration is produced. This current is monitored by a low leakage amplifier, giving a voltage readout for THC. To measure CH the sample gas is passed through the selective catalytic combustion unit (the NMHC cutter), which oxidizes NMHC without oxidizing Ok This is shown as A below. 8 represents the THC concentration measured without passing the gas through the NMHC cutter. Thus B- A will give the concentration of NMHC. The final concentration value is calculated using a relative-sensitivity correction coefficient, k, as shown below.

  • CH4 Concentration A
  • NMHC Concentration k (B - A)
  • THC Concentration A + k (B - A)

  • Principle: Flame ionization detection (FID) with selective combustion
  • Application: THC, NMHC, and CH4 in ambient air
  • Range: Standard ranges: 0-5/10/20/50 ppmC; auto range ~ manual range selectable; can be operated by remote switching.
    Optional (measurable) ranges: 4 ranges selectable from 0-100 ppmC, within 10 times range ratio;
    auto range ~ manual range selectable; can be operated by remote switching.
  • Repeatability: ±1.0% of F.S.
  • Linearity: ±1.0% of F.S.
  • Sample gas flow rate: Approx. 0.9 L/min
  • Indication: Measured value, range, alarm, maintenance screen
  • Alarms: During AIC, zero calibration error, span calibration error, temperature error in zero gas purifier, ignition failure error, etc.
  • On-screen messages are available in four languages: English, German, French, and Japanese.
  • Input/output:
    • 0-1 V/0-10 V/4-20 mA, to be specified (2 systems: either (1) momentary value and integrated or (2) moving average value)
    • Contact input/output
    • RS-232C
  • Power: 100/110/115/120/220/230/240 VAC, 50/60 Hz (to be specified)
  • Dimensions: 430(W) x 550(D) x 221(H) mm
  • Mass: Approx. 33 kg,
  • Notes: ppmC is shown as symbol, not as unit.