LumaSense Technologies, Inc.

- Model IGA 6/23 Advanced - Stationary, Digital Pyrometer for Non-contact Temperature Measurement

From

0

The pyrometer IGA 6/23 Advanced is a digital, compact, and fast infrared measuring instruments for non-contact temperature measurement on metals, ceramics, or graphite. For optimal match to the application, the instrument is equipped with a high-end optics with manual focus. The fast response time of only 0.5 ms facilitates the measurement of fast and dynamic processes or short temperature peaks. The integrated 4 digit LED display indicates the current measuring temperature or the currently set measuring distance. For a precise alignment of the pyrometers to the measuring object, the instruments are optionally equipped with a laser targeting light or a view finder. The pyrometers can be connected to a PC through an RS485 to USB connection, enabling you to make parameter adjustments using the InfraWin software. The software can be used for temperature indication, data logging, and further analyzing of complete temperature processes.

  • Wide temperature ranges for flexible process adaptation
  • Highest accuracy and repeatability in its class
  • Fully digital core for sub-ranging and adopted analog output
  • Response time of 0.5 ms for very fast and highly dynamic processes
  • High-end optics with manual focus capability
  • 4 digit LED display
  • Robust, stainless steel sensor for harsh environments (IP65/NEMA4)

  • Induction processes (e.g. Hardening, Welding, Brazing,
  • Soldering etc.)
  • Heating and cooling processes
  • Melting 
  • Casting
  • Annealing
  • Rolling
  • Forging
  • Sintering

  • Measurement Specifications
    • Temperature Ranges :
      • (MB 10) 50 to 1000 °C
      • (MB 13) 75 to 1300 °C
      • (MB 18) 150 to 1800 °C
    • Sub Range: Any range adjustable within the temperature range, minimum span 50 °C
    • Spectral Ranges: 2 to 2.6 μm (main wavelength 2.3 μm)
    • Resolution: 0.1 °C or 0.2 °F at interface; < 0.0015% of adjusted temperature range at analog output, 16 bit; 1 °C or 1 °F on display
    • Emissivity ε: 0.050 to 1.000 in steps of 1/1000
    • Transmittance τ: 0.050 to 1.000 in steps of 1/1000
    • Exposure Time t90: 0.5 ms; (with dynamic adaption at low signal levels) adjustable to min; 1 ms; 3 ms; 5 ms; 10 ms; 50 ms; 250 ms; 1 s; 3 s; 10 s
    • Measurement Uncertainty: (ε = 1, t90 = 1 s, TAmb. = 25 °C) < 1500 °C: 0.3% of reading in °C + 2 °C > 1500 °C: 0.6% of reading in °C
    • Repeatability: (ε = 1, t90 = 1 s, TAmb. = 25 °C) 0.15% of reading in °C + 1 °C
  • Optical Specifications
    • Sighting Built-in laser aiming light (max. power level < 1 mW, λ = 630 to 680 nm, CDRH class II) or through-lens sighting
    • Optics: Manually focusable from rear cover measuring distance a = 210 to 5000 mm
    • Distance Ratio:
    • MB 10: approx. 50 : 1
    • MB 13: approx. 100:1
    • MB 18: approx. 350:1
  • Environmental Specifications
    • Protection Class: IP 65 IEC 60529 (value in mated condition)
    • Operating Position: any
    • Ambient Temperature: 0 to 70 °C at housing
    • Storage Temperature -20 to 80 °C
    • Relative Humidity: Non condensating conditions
    • Weight: 0.6 kg
    • Housing: Stainless steel
    • CE Label: According to EU directives about electromagnetic immunity
  • Interface 
    • Connection: 12-pin connector
    • Display (in rear cover): LED, 4 digit matrix, 5 mm high temperature signal or measuring distance
    • Parameters: Adjustable via interface:
    • emissivity, sub range, ambient temperature compensation, settings for maximum value storage, address, baud rate, transmittance, response time t90, 0 to 20 mA or 4 to 20 mA analog output range, °C / °F
    • Readable via interface: measured value, internal temperature of the unit, measuring distance
  • Communication 
    • Analog Output: Adjustable 0 to 20 mA or 4 to
    • 20 mA, linear (via digital interface)
    • Digital Interface: RS485 addressable (half-duplex)
    • Baud rate: 1200 Bd to 115.2 kBd
    • (on request RS232 (not addressable))
    • Maximum Value Storage: Built-in single or double storage. Clearing with adjusted time tclear (off; 10 ms; 50 ms; 250 ms; 1 s; 5 s; 25 s), via interface, automatically with the next measuring object, hold-function
  • Electrical 
    • Power Supply: 24 V DC ± 25%, ripple must be less than 50 mV
    • Power Consumption: Max. 3 W ( incl. laser)
    • Load (analog output): 0 to 500 Ω
    • Isolation: Power supply, analog output and digital interface are galvanically isolated from each other

Customer reviews