Stationary, digital ratio pyrometer for non-contact temperature measurement in ranges between 600 and 3000 °C. The ISR 6 Advanced is a digital, compact and fast 2-color pyrometer for non-contact temperature measurement. The pyrometer measures in the 2-color method (ratio method) in which two adjacent wavelengths are used for the temperature determination.This technique offers the following advantages compared to standard 1-color pyrometers: the temperature measurement is largely independent of the object’s emissivity and in wide ranges unaffected by dust and other contaminants in the field of view.

The measuring object can be smaller than the spot size, measurements through dirty viewing windows are possible up to a certain contamination. Additionally, the pyrometer can be switched to 1-color mode and used like a conventional pyrometer in a spectral range near 0.9 μm. The response time of only 2 ms facilitates the measurement of fast processes.

The ISR 6 can be connected to a PC through an RS485 to USB connection, enabling parameter adjustments made using the InfraWin software. This can be used for temperature indication, data logging and further analyzing of complete temperature processes.

Optional Integrated Line Optics:
Besides the standard optical heads, the ISR 6 Advanced is also available with integrated line optics, which features a special spot in the shape of a line. This provides additional advantages for some applications such as wire production or pouring stream measurements.

  • Widest temperature ranges for most flexible process adaptation
  • Highest accuracy and repeatability in its class
  • “Dirty Window” Warning
  • Fully digital core for sub-ranging and adopted analog output
  • Very fast 2 ms response time for highly dynamic processes
  • Best optics in its class with manual focus capability
  • 4 digit LED display
  • Robust, stainless steel sensor for harsh environments (IP65/NEMA4)

  • Steel Making
  • Metal Processing - Induction Processes: Hardening, Welding, Forging, Brazing, Soldering, etc.
  • Metal Processing - Noble Metals Melting and Purifying
  • Metal Processing - Wire/Rod Mill Water Box Measurement, Laying Head & Air Cooling Conv.
  • Solar Industry - Silicon Processing Polycrystalline Cashing in Vacuum Melting Furnace, Silicon Ingot Growth in CVD Reactors (Siemens Process), Crystal Pulling of Monocrystalline Silicon (Czochralski Process)
  • Glass Industry - Gob Temperature Measurement
  • Cement Industry - Clinker temperature in rotary kilns

Measurement Specifications

  • Temperature Ranges:
    • (MB 14) 600 to 1400 °C
    • (MB 18) 700 to 1800 °C
    • (MB 25) 800 to 2500 °C
    • (MB 30) 1000 to 3000 °C
  • Sub Range: Any range adjustable within the temperature range, minimum span 50 °C
  • Spectral Ranges: Channel 1: 0.9 μm;
  • Channel 2: 1.05 μm
  • Resolution: 0.1 °C or 0.2 °F at interface; < 0.0015% of selected sub range at analog output, min. 0.1°C, 16 bit; 1°C or 1 °F on display
  • Emissivity ε: 0.050 to 1.000 in steps of 1/1000 (1-color mode)
  • Transmittance τ: 0.050 to 1.000 in steps of 1/1000 (1-color mode)
  • Emissivity Slope K: 0.800 to 1.200 in steps of 1/1000 (2-color mode)
  • Measurement Uncertainty: (ε = 1, t90 = 1 s, TAmb. = 25 °C): < 1500 °C: 0.3% of reading in °C + 2 °C > 1500 °C: 0.6% of reading in °C
  • Repeatability: (ε = 1, t90 = 1 s, TAmb. = 25 °C): 0.15% of reading in °C + 1 °C

Optical Specifications

  • Sighting: Built-in laser aiming light (max. power level < 1 mW, λ = 630 to 680 nm, CDRH class II) or through-lens sighting
  • Optics: Manually focusable from rear cover measuring distance a = 210 to 5000 mm
  • Distance Ratio:
    • MB 14 approx. 100 : 1
    • MB 18 approx. 190 : 1
    • MB 25 and MB 30 approx. 350 : 1

Environmental Specificatons

  • Protection Class: IP 65 IEC 60529 (value in mated condition)
  • Operating Position: any
  • Ambient Temperature: 0 to 65°C at housing
  • Storage Temperature: -20 to 80°C
  • Relative Humidity: Non condensating conditions
  • Weight: 0.6 kg
  • Housing: Stainless steel
  • CE Label: According to EU directives about electromagnetic immunity

Note: MB is a shortcut used for temperature range (in German: Messbereich)

Interface:

  • Connection: 12-pin connector
  • Display (in rear cover): LED, 4 digit matrix, 5 mm high for 2-color or 1-color temperature signal or measuring distance
  • Parameters: Adjustable via interface:
  • 2-color / 1-color temperature signal, accordingly emissivity slope or emissivity, temperature sub range, settings for maximum value storage, address, baud rate, switch off limit, 'dirty window' warning, transmittance, response time t90, 0 to 20 mA or 4 to 20 mA analog output range, °C / °F
  • Readable via interface: measured value, internal temperature of the unit, measuring distance

Communication:

  • Analog Output: Adjustable 0 to 20 mA or 4 to 20 mA, linear (via digital interface)
  • Digital Interface: RS485 addressable (half-duplex) Baud rate: 1200 Bd to 115.2 kBd (on request RS232 (not addressable)
  • Switch Off Limit: 2% to 50% (adjustable via interface)
  • 'Dirty Window' Warning: : Relay contact, max. continuous current 0.4 A, setting of the warning level: 0 (off) to 99%
  • Response Time t90: 2 ms (with dynamic adaption at low signal levels); adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
  • Maximum Value Storage: Built-in single or double storage. Clearing with adjusted time tclear (off; 0.01s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), via interface, automatically with the next measuring object, external contact, hold-function

Electrical:

  • Power Supply: 24 V DC ± 25%, ripple must be less than 50 mV
  • Power Consumption: Max. 3 W ( incl. laser)
  • Load (analog output): 0 to 500 Ω
  • Isolation: Power supply, analog output and digital interface are galvanically isolated from each other