Metrohm AG

- Model A629281130 - NIRS Analyzer Pro - Single Point Process Analysis System

From

0

The NIRS Analyzer Pro is a single point process analysis system based on high resolution diode array technology. It provides non-destructive analysis of solids like granules and powders. This system is typically mounted above a conveyor belt in order to analyze products passing by. Fast moving, uneven surfaces are able to be analyzed easily with this configuration.

  • Ambient Temperature : -5 - 40 °C
  • Dimensions in mm (W/H/D) : 420/420/135
  • Ingress Protection : IP69K
  • Number of sample streams : 1
  • Weight kg : 25

NIRS Analyzer Pro

  • Analysis time: 5 - 50 ms / integration time; average time per result: 1 - 15 seconds.
  • Wavelength range: 1100 - 1650 nm
  • Detector: InGaAs diode spectrum
  • Spectral dispersion: 1.1 nm / pixel

General

  • Lifetime of the light source: Dual lamp system MTBF = 17,500 h
  • Software package (sold separately): Vision for instrument operation and method development
  • Wavelength accuracy: 0.5 nm
  • Wavelength precision: < 0.02 nm
  • Wavelength stability: < 0.01 nm/°C
  • Noise: < 60 Micro-AU
  • Random vibrations: 0.4 Grms at 10 - 150 Hz in accordance with IEC 60068-2-64 / 0.4 Grms at 10 - 1250 Hz in accordance with international Metrohm standard (additional information available on request)
  • Temperature: -5 - 40 °C (23 - 104 °F). With purge -5 - 65 °C (23 - 149 °F).
  • Purge: Flow rate min. 5 l/min, > 99.99% anhydrous, > 99.9% without oil and fine particles up to 0.3 μm
  • Humidity: 10 - 90% relative
  • Dimensions: (W × H × D) 42 × 42 × 13.5 cm (16.5 × 16.5 × 5.3 inches) + holders for the device
  • Weight: 25 kg / 55 lbs
  • Cabinet: 1.5 mm (lid 2.5 mm) stainless steel EN 1.4301 (SS2333)
  • Protection: IP69K1) in accordance with IEC 60529 and DIN 40050 Part 9, NT ELEC 023
  • Power supply: disconnected from network or network-dependent (recommended) / 100 - 240 VAC, 50 - 60 Hz, 2.0 A, 150 W

1) IP6x is the highest protection against dust entering the device. IPx9K means protected against the effect of high-pressure water and/or steam cleaning at high temperature.

Process Communication Options

  • OPC
  • Digital protocols
  • MODBUS, PROFIBUS, etc.
  • VISION I/O Modules
  • Available with 8, 16, or 24 Analog Channels, 24 Digital Channels