- Model Ultrapen™ PT2 - pH Pen & Temperature Pen

From

0

This instrument is designed to be extremely accurate, fast and simple to use in diverse water quality applications. Advanced features include automatic temperature compensation; highly stable microprocessor-based circuitry; user-intuitive design; and waterproof housing. A true one-handed instrument, the PT2 is easy to calibrate and easy to use. To take a measurement, you simply push a button then dip the pen in solution. Results display in seconds.

  • Accuracy of ±0.01 pH
  • 1, 2 and 3 Point Calibration
  • Proprietary Sensor Design
  • Automatic temperature compensation
  • Reliable Repeatable Results
  • Rugged, reliable, shock-resistant fully encapsulated electronics
  • Durable body constructed of aircraft aluminum
  • Simple one button function
  • High capacity N-type battery included
  • Waterproof

  • Agriculture
  • Hydroponics
  • Desalination
  • Aquaculture
  • Pharmaceutical
  • Boilers and Cooling Towers
  • Fountain Solutions for Printing
  • Horticulture
  • Wastewater
  • Pool and Spa
  • Food and Beverage
  • Irrigation
  • Reverse Osmosis
  • Plus Many More!

pH

  • Range: 0.00-14.00 pH
  • Accuracy: ± 0.01 pH (After Wet Calibration)
  • Repeatability: ± 0.01
  • Resolution: 0.01 pH

Temperature

  • Range:0 - 71° C / 32 - 160° F
  • Accuracy: ± 0.1 ºC / ± 0.1 ºF
  • Repeatability: ± 0.1 ºC / ± 0.1 ºF
  • Resolution: 0.1 ºC / 0.1 ºF
  • Time to Reading Stabilization: 10-30 seconds
  • Power Consumption: Active Mode 37 mA, Sleep Mode 2 µA
  • Temperature Compensation: Automatic to 25ºC
  • Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
  • Weight: 54 g/1.9 oz. (without soaker cap and lanyard)
  • Case Material: Anodized Aircraft Aluminum with Protective Coating
  • Battery Type: N type, Alkaline, 1.5V
  • Calibration Solutions: pH 4, pH 7, pH 10
  • Operating/Storage Temperature: 0 - 55ºC or 32 - 131ºF
  • Enclosure Ratings: IP67 and NEMA 6
  • EN61236-1: 2006 – Annex A: 2008: Electrostatic discharge to case of instrument may cause PT2 to spontaneously power on. If this happens, the PT2 will power itself off within seconds.

Standard Solutions for calibration:
The ULTRAPEN PT2 requires pH 4, pH 7, and pH 10 buffer solutions for wet calibration and pH Sensor Storage Solution for proper storage.
Order model#s: PH4, PH7, PH10, and PHSS

Replacement pH Sensor:
The ULTRAPEN PT2 features a user replaceable pH sensor. Installation instructions included with sensor.
Order model#: RPT2

Customer reviews

No reviews were found for Myron L - Model Ultrapen™ PT2 - pH Pen & Temperature Pen. Be the first to review!