Turner Designs Hydrocarbon Instruments, Inc.

- Model TD-4100 XDC-GP - General Purpose Oil in Water Monitor

The TD-4100 XDC General Purpose oil in water monitor is a lower cost version of the industry standard TD-4100 XDC and is designed for General Purpose areas (non-explosive atmospheres). It was designed for normally clean water applications or dirty water applications where fouling can be controlled by chemical injection or optional on-board cleaning systems. It utilizes a clear tube flow cell to contain the sample water at pressure to prevent oxygen contamination in steam systems and to return the sample to the process at pressure. It uses a unique internal Cell Condition Monitor to alert operations of the need for cleaning or to trigger the optional on-board cleaning system. As with all Turner monitors, it includes a field validation method.

  • Fluorescence measurement technology with lowest detection limits available
  • Minimal sample conditioning plumbing hardware reduces cost
  • Easy calibration with long term stability
  • Low maintenance
  • User friendly, E09 Electronics Package
  • Ruggedized for harsh environments, non-hazardous areas
  • Custom configurations and wetted materials

  • Applications: Steam Condensate, Boiler Feed, Normally Clean Water, Cooling Water, Intake Protection and Dirty Water with Chemical cleaning if needed
  • Hydrocarbons: Fuel Oil, Crude Oil, Other Fuels and Lubricating Oils, Phenol, Styrene, Most Heat Transfer Fluids, Aromatic Chemicals, BTEX
  • Detection Range: Fuel Oil, Crude Oil 5 ppb - 500 ppm in steam condensate
  • Low Detection Limit: Most Other Hydrocarbons <50 ppb
  • Dimensions: Incl. Wall Mounting: 25” H x 44” W x 12.5” D (635 mm x 1118 mm x 318 mm), Incl. Floor Mounting: 68.4” H x 25.7” W x 25.7” D (1737 mm x 653 mm x 653 mm)
  • Weight: Floor Mount: 173 lbs (78.5 kg)
  • Local Display: Yes, PPM, PPB, or Raw Signal
  • Controls: External 316 SS Touch Pad
  • Power Requirement: 90 - 240 VAC, 50/60 Hz +/- 10%, 184 W, 1 ph, (24 VDC Optional)
  • Plumbing Requirements: Feed: 3/8' or 1/2' Tube, Return: 3/8'or 1/2' Tube
  • Sample Pressure: 100 psig (689 kPag) (For higher pressures consult factory)
  • Sample Temperature: 140 °F (60 °C) (For higher temperatures consult factory)
  • Ambient Temperature: -4 to 131 ˚F (-20 to 55 ˚C) (Heater or Vortex Cooler Optional)
  • Required Inst. Air Supply General Purpose: 10 - 20 SCFH (0.28 - 0.57 m3/hr)
  • Flow Cell Characteristic: Closed Tube with Internal Cell Condition Monitor
  • Pressurized Sample Return: Yes
  • Operational Principle: UV Fluorescence with application specific optical configurations
  • Calibration Stability: +/- 10% over 6 Months or better
  • Response Time: 3 seconds full scale spike change, continuous reading
  • Calibration: Multiple-point linear or non-linear, or uncalibrated. Holds two calibrations (one for each sample steam).
  • Field Validation: Via factory supplied water soluble dye or site prepared standard with hydrocarbons
  • Reagents: (For dirty water applications, a cleaning solution may be required)
  • Alarms: Early Warning / High Alarm / Cell Condition / Cleaning In Process
  • Alarm Contacts: Four user-settable, independently-protected dry contact relays
  • Communication: Standard E09 features with 4-20 mA isolated, 500 ohm impedance, selectable between Loop Powered and Instrument Powered, Ethernet. Optional HART, ModBus
  • Diagnostics: Self Diagnostics, internal failure linked to alarm, Cell Condition
  • Security: Password protects software, Lockable Enclosure 316 Stainless Steel
  • IP Rating: Electronics Cabinet 316 stainless steel, NEMA 4X, IP 66
  • Maintenance: Calibration check, clean flow cell
  • Hazardous Area Options: Use TD-4100 XDC with proper rating
  • Sample Pump: Optional
  • Cleaning System: Optional
  • Data Logger: Standard - see E09 Data Sheet
  • ISO Certified: 9001: 2008 Quality System

Not for use in explosive atmospheres

  • Steam condensate
  • Boiler feed
  • Normally clean water
  • Cooling water
  • Intake protection
  • Dirty water with chemical cleaning if needed

Hydrocarbons Detected

  • Fuel oil
  • Crude oil
  • Other fuels and lubricating oils
  • Phenol
  • Styrene
  • Most heat transfer fluids
  • Aromatic chemicals
  • BTEX

Detection Range

  • Fuel oil, crude oil 5 ppb-500 ppm in steam condensate
  • Most other hydrocarbons