Thermo Fisher Scientific - Model 146i - Multi-Gas Calibrator
From Thermo Fisher Scientific - Ambient Gas Analyzers
The Model 146i mass flow controller ozone generator, permeation tube oven, power supply and solenoid valves are integrated into a single microprocessor controlled unit. This permits easy to use menu driven software and a consistent set of operation screens for all calibration procedures. Additionally, the microprocessor can make many of the necessary calculations, thereby freeing the operator from having to make the calculations in the field or laboratory. If desired, the Model 146i can be operated remotely by a data logger or perform multi point calibrations.
The Thermo Scientific Model 146i Dynamic Gas Calib...
The Thermo Scientific Model 146i Dynamic Gas Calibrator supplies precise levels of ozone , carbon monoxide, non-methane hydrocarbons, sulfur dioxide, nitric oxide and nitrogen dioxide or other gases that a user may require. The gas levels are used to calibrate instruments that perform zero, precision and level 1 span checks, audits and multi point measurements. The design of the Model 146i meets or exceeds all published U.S. Environmental Protection Agency requirements for multipoint calibration, audit level 1 and 2 span and precision checks.
Key Features
- Ethernet connectivity for efficient remote access
- Enhanced user interface with one button programing and large display screen
- Flash memory for increased data storage and user downloadable software
- Enhanced electronic design optimizes product commanality
- Improved layout for easier accessibility to components
- Gas Phase Titration for Generation of O3 and NO2
- Permeation Oven for generation of SO2, NO2, H2S, NH3 and other gases.
Specifications:
Dilution System :
- Flow Measurement Accuracy : +/- 2% setpoint or +/- 1% FS, whichever is less from 20-100%FS
- Repeatabiltiy of Flow Control : +/- 0.2% FS (Porter floe controller spec- better than 146C)
- Linearity of Mass Flow Measurements : +/- 0.5% FS
- Flow Range of Dilution Air : 0-10 SLPM
- Optional Ranges : 0-5/0-20 SLPM
- Flow Range of Cylinder Gases : 0-100 SCCM
- Optional Ranges : 0-50/0-200 SCCM
- Zero Air Requirements : 10 SLPM @ 30PSI
- Optional Ranges : 20 SLPM @ 30PSI
- Calibration Gas Input Ports : 3, optional 6
- Diluent Gas Input Ports : 1
- Response Time : < 60 sec. to 99% (146C spec)
Ozone Generator Option :
- Maximum Output : 1ppm @ 6 SLPM
Minimum Output : 10ppb @ 6 SLPM
Physical Specifications :
- Test Channel Analog : 6 @ +/- 100 mv, 1,5,10 volts (user selectable
- Digital Control Outputs : 10 relay and 8 24vdc solenoid
- Digital control Inputs : 16
- Temperature Range : 0-40 deg C
- Weight : 51 lbs. (58 lbs for 220-240VAC)
- Dimensions : 16.75' W x 8.62' H x 23'D
- Power : 100VAC 50/60Hz, 115VAC 50/60Hz, 220-240VAC 50/60Hz, 275 Watts (with all options)