DOD Technologies Inc

- Model Select - Portable Benzene Gas Detector

From

0

The Tiger Select measures benzene selectively by use of a special sensor and a pre-filter tube in front of the PID to remove interfering chemicals present in fuel vapors.  A special mode is provided to measure total aromatics on a continuous basis. The Tiger Select Basic Instrument with 10.0eV Lamp includes: Li-ion Rechargable Battery, Battery Charger & Outlet Adaptor, Rubber Boot, Explorer Case (space for 44L or 116L Ecosmart Cylinder), USB Memory Stick, Draeger Benzene Pre-Filter Tubes & Tube Opener, Tiger Select Manual & Quick Start Guide, Carbon Filter Assembly & Adaptor, PTFE FIlter Discs, Removable Tool for MiniPID Stack, Sample Cleaning Compound, Lanyard & Sensor Seal).

Minimum Resolution
Select PPM

  • Standard running mode 0.1 ppm
  • TAC 0.01 ppm
  • Tube Mode 0.001 ppm

Minimum Resolution
Select PPB

  • Standard running mode 0.001 ppm
  • TAC 0.001 ppm
  • Tube Mode 0.001 ppm

Maximum Reading

  • Standard mode 20,000 ppm or 20,000 mg/m3 (gas dependent)
  • Tube mode 200 ppm or 639 mg/m3 benzene

Response Time

  • 120 seconds. Progressive indication of benzene breakthrough is displayed in real time.

Accuracy

  • 10% display reading one digit benzene

Linearity

  • 5% display reading one digit

Intrinsically Safe Approvals

  • II 1G Ex ia IIC T4 Ga
  • Tamb = -15c < Ta < +45c (With lithium ion battery pack)
  • Tamb = -15c < Ta < +45c (With alkaline battery pack)
  • ITS09ATEX26890X IECEx ITS 10.0036X
  • 3193491 conforms to UL Std. 913, 61010-1
  • Certified to CAN/CSA Std. C22.2 No. 61010-1

Battery Life

  • Li-ion: Life up to 24 hours, charge time 6.5 hours
  • Alkaline: 3 x AA, typically 8.5 hours of life

Lamps

  • 10.0 Ev Krypton PID lamp

Data Logging

  • > 120,000 data log points including date and time stamp

Communication

  • Direct USB 1.1

Calibration

  • 2 and 3 point calibration (via calibration kit accessory)

Alarm

  • Flashing LED’s amber (low alarm, Red (high alarm)
  • Sounder 95 dBA at 300 mm (12″) vibration on alarm
  • Pre-programmed TWA and STEL

Flow Rate

  • > 220 ml/min

Temperature

  • Operating -20 to 60 C, -4 to 140 F (non-intrinsically safe)
  • Humidity 0-99% RH (non-condensing)

Protection

  • Designed to IP65 (Heavy Rain)
  • 1180
  • EMC tested to EN61326-1:2006, EN50270:2006 & CFR 47:2008 Class A

Weight & Dimensions

  • Instrument (probe fitted no tube attached)
  • H x W x L: 18.3 x 3.5 x 2.4in
  • Weight: .745 kg
  • Packed Weight: 5 kg

  • Weight: 11 lbs
  • Dimensions: 18.3 × 3.5 × 2.4 in
  • Units: Tiger Select – PPM /H&S/ Data Logging, Tiger Select – PPB /H&S/ Data Logging