Zwinsoft - Model ZWIN-AQMS30 - Ambient Air Quality Automatic Monitoring System
The ZWIN-AQMS30 ambient air quality automatic monitoring system can realize the timely grasp of the current situation of air quality in a certain area, and form a complete network for real-time sampling and analysis of ambient air in an area. It is mainly used in various environmental protection departments, electric power, petroleum, Monitoring of ambient air quality by large industrial and mining enterprises in chemical, steel, metallurgy, building materials, and Traffic environmental pollution monitoring, etc.
Products Details
The ZWIN-AQMS30 ambient air quality automatic monitoring system developed by our company is the latest high-tech precision instrument and equipment developed by using the international advanced photoelectric technology to realize continuous real-time monitoring of PM 10 , PM 2.5 , SO 2, NO 2 , O 3 , CO in ambient air ,and other pollutant concentrations, as well as monitoring of wind speed, wind direction, temperature, relative humidity, atmospheric pressure, etc.
The system adopts RS485 communication interface. RS485 can communicate with multiple computers, and the transmission distance is longer. The standard value output of this system is 0-5V, 4-20mA and has RS485 two-way communication control interface. All instruments have good anti-interference ability and self-recovery function after power failure, and the effective data capture rate is higher than 90%.
The monitoring system adopts a modular structure, with reasonable design and simple operation. It can realize remote data transmission and fault diagnosis.
Executive Standard
- Technical Specifications for Installation and Acceptance of Continuous Automatic Monitoring System for Ambient Air Gaseous Pollutants (SO2, NO2, O3, CO) ” ( HJ193-2013 )
- “Ambient Air Quality Standard” (GB3095-2012)
- “Technical Requirements and Detection Methods of Continuous Automatic Monitoring System for Ambient Air Particulate Matter (PM10, PM2.5)” (HJ653-2013)
- “Technical requirements and detection methods of continuous automatic monitoring system for ambient air gaseous pollutants (SO2, NO2, CO, O3)” (HJ654-2013)
- ” Technical Specification for Installation and Acceptance of Continuous Automatic Monitoring System for Ambient Air Particulate Matter (PM10 and PM2.5) ” ( HJ655-2013 )
- “Data Transmission Standard of Pollution Source Online Automatic Monitoring (Monitoring) System” (HJ/T212-2005)
- “Technical Specifications for Automatic Monitoring Information Transmission and Exchange of Environmental Pollution Sources (Trial)” (HJ/T 352-2007)
Note :Instruments Follow The Environmental Protection Standards of People’s Republic of China
System Structure
The whole system is mainly composed of analytical instruments, gas calibration system, data acquisition system, sampling system, meteorological monitoring system, communication and station building and other subsystems.
Pollution analysis instruments : including inhalable particulate matter (PM10) monitors, fine particulate matter (PM2.5) monitors, sulfur dioxide (SO2) analyzers, nitrogen oxides (NO x ) analyzers, ozone ( O3 ) analyzers, carbon monoxide ( CO ) analyzer, etc .;
Gas calibration : including zero gas generator, dynamic gas calibrator, standard gas and accessories;
Data acquisition system: The system platform can view real-time and historical measurement data, organize and store data and export to other formats, and form complete reports such as curves and reports; remotely monitor system parameters, collect analyzer event record data, diagnose menus and initialize calibration ;
Sampling system: gas sampling main pipe, multi-branch joint, exhaust fan ;
Meteorological monitor : temperature, relative humidity, air pressure, wind speed, wind direction (optional rainfall, sunshine, radiation, etc.) ;
Communication mode : with serial communication ( RS232/RS485 ), wireless communication ( GPRS/CDMA/ ), wired communication ;
Station building : steel structure foundation/station main body /combination cabinet/indoor temperature and humidity/air conditioner/ventilation fan/uninterruptible power supply/fire alarm/firefighting facilities, etc.
Sampling System
- The sampling head should be able to prevent rainwater, coarse particles and insects from entering the main pipe;
- The sampling main pipe is a multi-branch waterproof sampling pipe, and the material should be selected from materials that do not chemically react with the monitored pollutants and do not release interfering substances, and have the function of heating and heat preservation;
- The inner diameter of the main pipe is selected between 1.5-15cm, the air flow in the sampling main pipe should be kept in a laminar flow state, and the residence time of the gas in the main pipe is less than 20 seconds;
- The number of branch pipes meets the needs of all gaseous projects;
- The length of the sampling pipe should be able to ensure that it is 1.2 meters higher than the roof of the station building (to ensure that the sampling is not affected by surrounding obstacles);
- The sampling system is sealed and connected with the house body with flanges or other types of multi-stage water-proof connections; the flanges connected with the house body must be corrosion-resistant and strong stainless steel;
- The main pipeline of the sampling system is detachable, which is convenient for disassembly, cleaning and maintenance without affecting the external flange connection of the roof and the connection of the instrument end .
Cabinet
It is used for the installation and normal operation of particulate matter instruments, gaseous analysis instruments, calibration instruments, data collectors, industrial computers, VPNs, etc., including corresponding other supporting equipment.
- 1 9-inch standard cabinet with good heat dissipation performance and guide rails for instrument installation and movement ;
- The cabinet adopts aviation-grade guide rails to connect and load instruments, and the cabinet partitions are drawer type, which can provide a reasonable installation layout according to the physical size of various types of instruments and equipment;
- The whole cabinet is designed with welded frame, using high-strength cold-rolled steel plate and heavy-duty integrated chassis, with a maximum bearing capacity of 800KG;
- Each cabinet is equipped with an appropriate amount of power strips, and the cabinet has grounding holes;
- Configure the sampling main pipe, sampling head, sampling fan and sampling branch pipe (Teflon) connected with the instrument for gas analysis equipment ;
- The cabinet is equipped with a vertical LCD display mounting plate and a keyboard support plate.
Power Supply
Can meet the needs of SO 2 , NO 2 , CO, O 3 , PM 2.5 , PM 10 analyzers, zero gas generators, calibrators, data collectors and other equipment, ensuring long-term stable operation of the above instruments and equipment, and is not affected by induced electricity. Variable voltage, the regulated power supply can load more than 5kW for the normal use of the instrument, and the regulated power supply is grounded.
- Nominal capacity: 3kVA
- Input voltage range: 130V~300V
- Input frequency range: 50Hz±0.5Hz
- Voltage regulation accuracy:
Input: 187V~250V, output 220V±2% (full load)
Input: 170V~280V, output ≤220V+3% (full load), output 220V -7% (full load)
Input: 130V~300V, output ≤220V+3% (half load), output 220V -7% (full load)
- Voltage adjustment time: ≤40ms
- Power factor: ≥0.95
- Efficiency: 85%-92%
- Audio noise: <53dB
- Insulation resistance: ≥2MΩ
- Dielectric strength: power frequency sinusoidal voltage 1500V for 1 minute without breakdown and arcing
- Ambient humidity: -25 ? ~+40 ?
- Relative humidity: <97%
- How it works: Continuous
Data Collector
- CPU: frequency above 3.0GHz or above 2.4GHz dual-core; memory: above 2G; hard disk: above 320G/7200R;
- Standard configuration: 16 or more RS232/RS485 communication ports, including no less than 8 independent serial ports;
- Chassis: 19-inch 4U industrial chassis (industrial power supply supports at least 300W load);
- Operating system: pre-installed windows 2012 server professional version or above;
- Keyboard and display: Universal 104-key keyboard, LCD display 1024*768 pixels or more;
- Two or more RJ45 ports;
- Interface expansion module: Select the following interface modules (RS232/ R485) depending on the configuration and integration of the site equipment
- At the same time, it is a compatible and unified network management platform ;
- Support plug and play of new devices , automatically identify device access mode and prompt users to configure and register the device;
- Able to identify and distinguish the data of the device in various states, such as normal data, calibration data, etc .;
- Through the relevant interface of the data acquisition module, you can browse and control the air quality monitoring equipment of various mainstream manufacturers in a simple and unified way. The supported operations include at least: instrument calibration, state setting, instrument restart or reset, etc.;
- Support reading, browsing, querying and data exporting of real-time data and historical data of devices
- Support strategy configuration (task customization) operation, and realize automatic execution of various functions without manual participation;
- Support mobile terminal to remotely operate equipment through 3G network, including: real-time data collection, equipment status browsing, instrument remote calibration, etc
Station Building
- The area of the station building is not less than 15 m 2 ( outside: 3000mm×5000mm×2650mm , inside: 2800mm×4800mm×2500mm ) The service life of the new station building is ≥10 years under normal external conditions and normal maintenance conditions. The ground of the station building shall be a platform base of C25 concrete.
- The station building has a flat roof with a slope of not more than 10 °. The main structure should be designed to be able to withstand severe weather. It should have the properties of heat insulation, airtightness, durability, fire resistance, shock resistance, etc.; there are waterproof and leak-proof measures. The load-bearing capacity of the top of the station building is greater than or equal to 250kg/m 2 .
- The station building has no window or double-layer sealed window structure , and the wall should have good thermal insulation performance.
- Temperature and humidity control equipment should be installed in the station building, so that the indoor temperature of the station building is 25 ? ± 5 ? , and the relative humidity is controlled at 80%.
- The station building should have waterproof and moisture-proof measures, and the stratum of the station building should be 25 cm away from the ground (or roof).
- The position of the exhaust port of the exhaust fan of the sampling device and the exhaust port of the monitoring instrument should be set on the wall near the lower part of the station building, and the distance between the exhaust port and the ground in the station building should be more than 20cm.
- When a meteorological pole or a meteorological tower for fixing the meteorological sensor is set on the roof of the station, the vertical height of the meteorological pole, tower and the station roof shall not be less than 2 meters, and the building structure of the meteorological pole, tower and station building shall be able to Withstand wind force of 10 or more.
- Install a protective fence on the roof of the station building, the height of the protective fence should not be less than 1.2 meters, and reserve the installation hole for the sampling main pipe. There are simple stairs outside the station building, which is convenient for daily maintenance.
- The power supply of the station building adopts three-phase five-wire power supply. The cable must be larger than the national standard cable of 10mm2 . The power distribution box is installed at the entrance, and three single-phase 15A air switches are installed in the power distribution box for the three-phase power supply. The main switch is used in separate phases; the wiring of the station building is required to be beautiful, and the wiring should be equipped with a trunking.
- The lighting system of the station building includes fluorescent lamp groups, wiring, switches, and sockets; lamps should be complete, installed firmly and reliably, insulating materials for fixing live parts of lamps and insulating materials for protection against electric shock, and should be resistant to combustion and open flames.