TAPI - Model T360M - Mid-Range Gas Filter Correlation CO2 Analyzer
From Carbon Compound Instruments
Using a gas filter correlation wheel, the Model T360M Mid-Range CO2 Analyzer measures carbon dioxide by comparing infrared energy absorbed by a sample to that absorbed by a reference gas according to the Beer-Lambert law. This design produces excellent zero and span stability and high signal-to-noise ratio, allowing excellent performance over a wide concentration range, making this instrument an ideal choice for CO2 reporting requirements associated with dilution CEMS.
-
Most popular related searches
Description
The energy loss through the sample cell is compared with the reference signal provided by the filter wheel to produce a signal proportional to concentration, with little effect from interfering gases within the sample. This design produces excellent zero and span stability and high signal to noise ratio, allowing excellent performance over a wide concentration range, making this instrument an ideal choice for CO2 reporting requirements associated with dilution CEMS.
All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.
The Model T360M comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.
Features:
- Ranges: 0-4 ppm to 0-4,000 ppm, user selectable
- Dual ranges
- Large, vivid, and durable color graphics display with touch screen interface
- Ethernet, RS-232, and (optional) USB com ports
- Front panel USB connections for peripheral devices and firmware upgrades
- 8 analog inputs (optional)
- Adaptive signal filtering optimizes response time
- Temperature & pressure compensation
- Comprehensive internal data logging with programmable averaging periods
- Ability to log virtually any operating parameter
- Two-year warranty
- Five-year warranty on GFC wheel
Specifications:
- Ranges :
- Min: 0 - 4 ppm full scale
- Max: 0 - 4,000 ppm full scale (selectable, dual-range supported)
- Measurement Units : ppb, ppm, μg/m3, mg/m3 (selectable)
- Zero Noise : < 0.2 ppb (RMS)
- Span Noise : < 1% of reading (RMS)
- Lower Detectable Limit : 0.4 ppm
- Zero Drift : < 0.5 ppm/24 hours
- Span Drift : < 0.5% of full scale/24 hours
- Lag Time : 10 seconds
- Rise/Fall Time : < 60 seconds to 95%
- Linearity : 1% of full scale
- Precision : 0.5% of reading
- Sample Flow Rate : 800 cc/min ±10%
- Power Requirements : 100V-120V, 220V-240V, 50/60 Hz
- Analog Output Ranges : 10V, 5V, 1V, 0.1V (selectable)
- Recorder Offset : ±10%
- Included I/O :
- 1 x Ethernet: 10/100Base-T
- 2 x RS232 (300-115,200 baud)
- 2 x USB device ports
- 8 x opto-isolated digital outputs
- 6 x opto-isolated digital inputs
- 4 x analog outputs
- Optional I/O :
- 1 x Ethernet: 10/100Base-T
- 2 x RS232 (300-115,200 baud)
- 2 x USB device ports
- 8 x opto-isolated digital outputs
- 6 x opto-isolated digital inputs
- 4 x analog outputs
- Operating Temperature Range : 5 - 40ºC
- Dimensions (HxWxD) : 7” x 17” x 23.5” (178 x 432 x 597 mm)
- Weight : 40 lbs (18.1 kg)
Customer reviews
No reviews were found for TAPI - Model T360M - Mid-Range Gas Filter Correlation CO2 Analyzer. Be the first to review!