How to Calibrate Sensistor ISH2000 Hydrogen Leak Detector - Video

0
Purpose: Calibrate Sensistor ISH2000 with a certified reference leak RL2000
Test steps: Connect leak to tracer gas source - Select 'Calibration' on Sensistor ISH2000 menu -- start calibration -- apply sensor to leak -- save calibration by pressing on hand probe button.
Instruments: Sensistor ISH2000 Hydrogen Leak Detector, Reference Leak RL2000.
Leak limit: This leak was certified at a flow of 5x10- 4 std cc/s.