Tianezk Instrument Manufacturing (Chengdu) Co., Ltd.
  1. Companies
  2. Tianezk Instrument Manufacturing ...
  3. Articles
  4. How to properly troubleshoot problems ...

How to properly troubleshoot problems with a vacuum parallel concentrator

SHARE
Oct. 16, 2025

The vacuum parallel concentrator is a core piece of equipment for sample pretreatment, and its malfunctions directly affect experimental efficiency and the accuracy of results. The following is a systematic overview of typical malfunctions, their causes, and targeted solutions:

1. Abnormal vacuum level (failed to reach the set value)
Characteristics: The instrument displays a low vacuum level and a slow evaporation rate.

Potential reasons:
Vacuum pump efficiency decline: Long-term operation leads to vane wear or oil accumulation, requiring disassembly and cleaning of the pump chamber and replacement of the drain oil; if the noise increases and the pumping speed decreases, the vacuum pump may need to be replaced.

Sealing failure: Observe whether there is any air leakage sound at each connection. Pay special attention to whether the sample holder sealing ring has hardened and cracked. When replacing the O-ring, apply an appropriate amount of vacuum grease to enhance the sealing performance.

Cold trap blockage: The accumulation of organic solvent vapor condensate obstructs airflow. Clean the inner wall of the cold trap with anhydrous ethanol, and replace the molecular sieve desiccant if necessary.

Pipeline blockage: Check if the vent valve is open, and use compressed air to back-purge the vacuum pipeline to remove particulate blockage.

II. Heating runaway (temperature deviation > ±5?)
Characteristics: The actual temperature far exceeds the set value, and the sample boils and splashes out.

Solution:
Temperature control module calibration: Measure the temperature of the aluminum block slots with a standard thermometer. If the deviation is too large, the PID parameters need to be recalibrated; replace the aged thermocouple probe.

Heating plate inspection: After disconnecting the power, check the resistance of the graphite heating plate. If there is a partial short circuit, the entire heating unit needs to be replaced; clean the surface of the heating plate to remove any crystallized salt residue.

Emergency braking: Immediately disconnect the power supply and allow it to cool naturally to below 60°C before opening the lid to prevent burns.

3. Program execution disorder (screen error/action interruption)

Typical code analysis:

  • E03 (Communication Timeout): Check if the RS485 data cable connector is loose, and restart the controller host; if it occurs frequently, the firmware version needs to be upgraded.
  • P07 (Abnormal Air Pressure): Confirm that the exhaust gas pipeline is unobstructed and adjust the pressure threshold parameter to a reasonable range (usually -90~-60kPa).
  • Emergency procedure: Press and hold the reset button for 10 seconds to restore factory settings and re-import the verified program file.

IV. Sample Cross-Contamination

root cause:
Volatile substances diffuse with the airflow to adjacent wells, especially during continuous processing of high-concentration samples.

Improvement measures:

  • Install baffles: Insert polytetrafluoroethylene baffles between sample holders to block the aerosol migration path.
  • Optimized process: A gradient pressure increase and vacuum release method is adopted. After each batch of samples is processed, nitrogen gas is used to ventilate and remove waste before opening the lid.
  • Cleaning: Use a special solvent to ultrasonically clean the glass sleeve, and bake at high temperature to remove stubborn residues.

V. Mechanical movement sluggishness

Diagnostic points:
Check if there are any foreign objects blocking the guide rail of the lifting mechanism. Lack of oil in the lead screw nut can cause abnormal noise during lifting. Inject a small amount of silicone-based grease into the linear bearing and adjust the belt tension to a pressure deformation of 1cm.

By establishing a comprehensive troubleshooting system that integrates "phenomenon-cause-countermeasure" and combining it with regular maintenance, the stable operation of equipment can be guaranteed to the greatest extent. When encountering complex faults, it is recommended to contact the original manufacturer`s engineers for in-depth diagnosis; do not blindly disassemble core components.

Most popular related searches