Diazyme - Model DZ170A - Human Lambda Free Light Chain Assay - Brochure

Diazyme`s Human Lambda Free Light Chain (FLC) assay has excellent across the Analytical Measuring Range (AMR) in standard mode: 6.1 – 200 mg/L ; extended mode: 6.1 – 4000 mg/L. The method comparison gave a linear regression coefficient of 0.972, a slope of 1.054 and y-intercept of 0.636 mg/L. Precision of the assay was evaluated according to the CLSI EP5-A guidelines and has a Within-Run precision from 1.1% - 6.5% and a Total Precision of 1.1% - 11.4%. The Lambda assay displays outstanding sensitivity with an LOD of 2.9 mg/L; LOQ of 6.1 mg/L; and a LOB determined to be 1.7 mg/L.