FEATURES
• TFT LCD 64k color display
• LED stabilized
• 12 channel, Sirf III GPS
• Reachback via Bluetooth© connected to DUN
capable cell phone
• ANSI N42.42 output format
• Web interface for monitoring and configuring
instrument
• Original three button operation
• 1 GB event data storage
• Visible, audible and tactile alarm annunciators
• Embedded Windows CE operating system
• Meets ANSI N42.34 shock conditions
NEXT GENERATION HANDHELD
RADIONUCLIDE IDENTIFICATION DEVICE
identiFINDER® 2
The identiFINDER® 2 is a logical enhancement of the
original identiFINDER series of handheld instruments for the
detection and identification of radionuclides. The owners and
operators of the 10,000+ identiFINDERs currently deployed
worldwide have provided ICx with a tremendous resource for
suggestions and comments on how to improve the already
exceptional identiFINDER. The identiFINDER 2 is the result of
these comments and suggestions.
The monochrome LCD display has been replaced with a TFT
LCD, 64k color, 320 by 240 pixel display that is readable in
virtually all light conditions. The spectrum is continuously
source peak or LED peak stabilized to handle a wide range
of count rates and conditions. The advantage of the LED
stabilization technique is that there is no peak interference
in the identification spectrum. Wired communication uses a
micro USB connector, USB 2.0, and wireless communication
is via a Class 2.0 Bluetooth© interface, with a 33 ft (10m)
range. Both communication methods support the reach-back
capability provided with the identiFINDER 2. A Web interface
is provided via USB emulated TCPIP (LAN) for monitoring
and configuring the identiFINDER 2. A twelve channel, Sirf III
GPS is included for incident location. Data storage has been
enlarged to 1 GB to save more spectra and event information.
Features of the original identiFINDER that are unchanged
are the size, shape and weight; the three button (fourth for
power) operation; the scintillation, neutron and GM detector
types, sizes and sensitivities; the screens and menus, when
operated in the original identiFINDER mode; the types of
alarms; the radionuclide identification accuracy; and the
reliability, support and service.
Annunciators and Displays
All components clearly marked:
1 12 channel GPS, SIRF III antennae
2 Display: Alarm visualization
3 Red LED (doserate alarm)
4 Blue LED (neutron alarm)
5 Tacticle alarm vibrator (inside housing)
6 powerPACK Ultra (inside removable base)
MEASURING
FINDING
ANALYZING
IDENTIFYING
identiFINDER® 2 Series
1
2
3
4
5
6
1 Green LED (external power)
2 Orange LED (charging)
3 Audible alarm speaker (inside housing)
4 USB 2.0, micro-B connector
5 Bluetooth© (inside housing)
identiFINDER® 2-ULCS-NG
• Internal 3 nCi Cs137 source included for start-up energy calibration verification
• LED peak stabilized with NaI gamma detector
• GM detector for high doserate situations
identiFINDER® 2-ULCS-NGH
• Internal 3 nCi Cs137 source included for start-up energy calibration verification
• LED peak stabilized with NaI gamma detector
• GM detector for high doserate situations
• Moderated He3 neutron detector
identiFINDER® 2-NG
• Internal 15 nCi CS137 source included for start-up energy calibration verification
• Spectral stabilization with NaI gamma detector
• GM detector for high doserate situations
identiFINDER® 2-ULK-NG
• External K40 (250 ± 10 grams of KCl) included for start-up energy calibration
verification
• LED peak stabilized with NaI gamma detector
• GM detector for high doserate situations
identiFINDER® 2-NGH
• Internal 15 nCi CS137 source included for start-up energy calibration verification
• Spectral stabilization with NaI gamma detector
• GM detector for high doserate situations
• Moderated He3 neutron detector
identiFINDER® 2-ULK-NGH
• External K40 (250 ± 10 grams of KCl) included for start-up energy calibration
verification
• LED peak stabilized with NaI gamma detector
• GM detector for high doserate situations
• Moderated He3 neutron detector
Six Available Models
DETECTOR TYPES
Gamma NaI, 1.4 in x 2 in (35 mm x 51 mm)
High Dose Gamma GM detector
Neutron Moderated He3 proportional detector
0.6 in x 2.1 in (15 mm x 54 mm) @ 8 atm
PERFORMANCE
Gamma Energy Range 20 keV to 3 MeV
Gamma Resolution = 8% @ 662 keV (typically)
Gamma Sensitivity > 900 cps/µSv/h (>9 cps/µR/h) for
Cs137, 662 keV
Doserate Range 0.01 µSv/h to 1 Sv/h (1 µR/h to 100 R/h)
Dose Range 0.01 µSv to 1 Sv (1 µR to 100 R)
Neutron Sensitivity 2.6 CPS/nv ± 20%
Stabilization Cs137 source peak or LED peak; ± 1% for
temperature change rate of <0.5°C/min
Nuclide Identification Per ANSI N42.34
OPERATING CONDITIONS
Temperature -4°F to +131°F (-20°C to +55°C)
Humidity 10% to 80%, non-condensing
Shock According to ANSI N42.34
Protection Rating IP53 according to IEC 60529
INPUTS / OUTPUTS
DC Power / Charger + 9V DC / 2A
Bluetooth© Class 2.0, max range 33 ft (10 m)
Activated on demand
USB USB 2.0, micro-B connector
GPS Built-in, 12 channel SIRF III receiver
Web Interface Via USB emulated TCPIP (LAN)
Data Format ANSI N42.42 2006; standard for DHS
PHYSICAL
Size 9.75 in x 3.67 in x 2.95 in
(248 mm x 93 mm x 75 mm)
Weight 2.65 lbs (1.2 kg) with neutron detector
2.6 lbs (1.18 kg) w/o neutron detector
Display TFT, True Color LCD, 64k colors
Display Size 2.7 in (68.4 mm) diagonal, 320 * 240 pixel
R
A
_I
D
EN
TI
FI
N
D
ER
2_
10
25
10
SPECIFICATIONS
The Americas
2100 Crystal Drive
Suite 650
Arlington, VA 22202
T +1. 877.692.2120
sales@icxt.com
Middle East
Suite 1-11
Building 6E-A
Dubai Airport Freezone
PO Box 371363
Dubai, UAE
T +971 4 701 7195
F +971 4 701 7194
mesales@icxt.com
Europe
Piepersberg 12
42653 Solingen
Germany
T +49 212 222090
F +49 212 201045
europesales@icxt.com
Asia
Level 28 Gateway East
152 Beach Road
Singapore
T +65.6827.9789
F +65.6295.2567
asiasales@icxt.com
POWER
Power
Operating Time
Charging Time
MEMORY
Memory
Data Storage Capacity
Data Logging
Screen Shots
Alarms
SOFTWARE
Embedded Software
Web Interface
Reachback
Stride Data Protocol
One powerPACK Ultra with 4 ea
NiMH AA @ 2200 mAh @ 1.2 V,
One powerPACK 2
> 8 hours, full charge (doserate
mode)
4 hours, full charge (fast charge)
1.8 GB
600,000 radionuclide
identifications including
spectrum
Every 5 secs for 5 years w/o
neutron data
Every 5 secs for 2.5 years w/
neutron data
= 2,500
= 2,500
Windows CE operating system
TCP/IP emulated via USB
ANSI N42.42 data via
Bluetooth© to DUN compatible
cell phone
Stride Data Protocol for network
integration
Most popular related searches