Multi-Parameter Analyzer
4 TO 8 SENSORS
The Insite Instrumentation Group MPA 48 multi-channel analyzer is a unique system that combines ad-
vanced electronics with the latest in sensor technology. The analyzer will accept any combination of DO,
SS, pH, or ORP sensors and automatically configure for the correct operation. The system allows for flexi-
ble and economical process monitoring and control. The DO sensor does not need membranes, fill solu-
tions, nor replaceable sensor caps. The SS sensor is based on near infrared technology, eliminating inaccu-
racies caused by changes in the process color. The pH electrodes are flat glass, double reference junction
Optical Dissolved Oxygen—Suspended Solids—pH—ORP—Any Sensor Combination
FLUORESCENCE DISSOLVED OXYGEN —THEORY OF OPERATION
A very specific energy wavelength is transmitted to a ruthenium compound immobilized in a sol-gel matrix. The ruthe-
nium will absorb this energy, changing the outer electron’s energy level. The electron will then collapse back to it’s
original energy state, emitting the energy as a photon with a different specific wavelength. This is called fluorescing. If
the intensity of the transmitted wavelength is tightly controlled, the amount of fluorescing is both predictable and re-
peatable. If oxygen molecules are present the amount of fluorescing is reduced, referred to as fluorescence quench-
ing. By measuring the amount of quenching it is possible to determine the amount of oxygen present.
MPA 48MPA 48MPA 48MPA 48
MULTIMULTIMULTIMULTI----PARAMETER ANALYZERPARAMETER ANALYZERPARAMETER ANALYZERPARAMETER ANALYZER
MPA 48 MULTIPLE CHANNEL
PROCESS ANALYZER
Insite Instrumentation Group
80 Whisperwood Blvd.,Suite 107
Slidell, LA 70458
Phone - 985-639-0006
Fax - 985-639-0014
e-mail - info@insiteig.com
Website– www.insiteig.com
with temperature compensation. The ORP elec-
trodes are flat glass, double reference junction.
Both pH and ORP electrodes feature a quick dis-
connect design that makes changing electrodes a
snap and reduces down time due to electrode main-
tenance.
Measuring Range DO — 0 to 25 ppm
SS — 0 to 30,000 mg/l
pH — 2 to 12 pH
ORP- +/- 2000mV
Accuracy DO — 1% of reading or .05 ppm, whichever is greater
SS — 3% of reading
Sensitivity / Resolution DO — .01 ppm below 4.00, .1 ppm above 4.0
SS — 1 mg/l below 1000 mg/l
10 mg/l between 1,000 and 9,999 mg/l
100 mg/l above 10,000 mg/l
Repeatability DO — .01 ppm
SS — +/- 1%
Sensor Drift DO & SS — Less than 1% per year
Temperature Range DO & SS — 0 to 60 degrees C
pH & ORP — 10 to 100 degrees C
Response Time 95% in less than 60 seconds
Sensor Check Automatic self diagnostics
Outputs Up to 8 Opto-isolated 4-20 milliamp (one for each channel)
Optically isolated RS-485 Modbus
RS-232 serial
Up to 8 dry contact 10 amp relay setpoints (one for each
channel)
1 alarm relay
1 clean relay
Memory Backup Yes
Display Backlit graphical LCD display with UV protection
Contrast adjustment via keypad
Continuously displays all channels
Sensor Cable Length 33 feet standard (optional lengths up to 2000 ft)
Ambient Temperature minus 20 degrees C to 70 degrees C
Ambient Humidity 0 to 100 percent
Enclosure Rating NEMA 4X
Wetted Materials Epoxy, polyurethane, glass, and PVC
Maximum Pressure 100 psi
MPA 48 SPECIFICATIONS
Insite Instrumentation Group
e-mail - info@insiteig.com
Website– www.insiteig.com
OR CONTACT YOUR LOCAL TECHNICAL REPRESENTATIVE AT: