Ducom Instruments (USA) Inc.
- Home
- Companies
- Ducom Instruments (USA) Inc.
- Products
- ABEL - Model AFP 1 - Flash Point Tester
ABEL - Model AFP 1 -Flash Point Tester
DETERMINE THE FLAMMABLE AND COMBUSTIBLE PROPERTIES OF SAMPLES WITH A FLASH POINT BETWEEN -18°C AND 71°C. The AFP offers fully-automatic, temperature-controlled flash point determination of solvents, kerosene and chemicals. It is designed with an ionization detector, an automated shutter system and electric ignitors.
Most popular related searches
flash point
flash point testing
flame detection system
digital temperature controller
panel PC
flame detection
high-temperature monitoring
temperature monitoring system
temperature monitoring
flash testing
MEASURING PARAMETERS
- Temperature: °C/°F
- Measuring Range:-50°C to +100°C
- Resolution: 0.06°C
- Accuracy: ±0.1°C
- Repeatability/Reproducibility: as per standard methods or better
- PT100, Class A
- Dimensions: 31 x 47 x 52 cm (WxDxH)
- Weight: 27 kg
- Power: 220V + 15% / 50 to 60 Hz (or) 115V + 15% / 60 Hz
- Integrated Touchscreen Panel PC: 8.4-inch, TFT/LCD, high-resolution display
- Shutter Mechanism: Automatic system opens and closes the shutter; complies with all methods and specifications
- User-Friendly Software Interface: automated, temperature-controlled testing
- Storage Capacity: Users can store data files in an Excel-compatible format
HEATING
- Electrical heater
- Equipped with thermal cut-off system
TEST CUP
- Brass cup
- High temperature resistant handle
- Dual ignition system
- Ionization detector
- Electrical system or flame detection device
EN AND IP METHODS
- EN 924: Solvent-borne and solvent-free adhesives: Determination of flash point
- IP 170: Determination of flash point - Abel closed-cut method
- IP491: Determination of flash/no flash - Closed-cup equilibrium method
- IP 492: Determination of flash point - Closed-cup equilibrium method
ISO METHODS
- ISO 1516: Determination of flash/no flash - Closed-cup equilibrium method
- ISO 3679: Determination of flash/no flash point - Rapid equilibrium closed-cup method
- ISO 13736: Determination of flash point - Abel closed-cup method