LumaSense Technologies, Inc., An Advanced Energy Company

- Model ANDROS 6552 - OEM Gas Analyzer

From

0

Advanced Energy’s Andros 6552 gas analyzer is available with two gas channels and can be applied to any type or combination of refrigeration systems that employ FreonsTM and carbon dioxide. Ability to measure up to five gases. Rapid warm-up. Highly accurate calibration. Simple and fast system integration. RS232 or USB communications.

The Andros® 6552 gas analyzer / non-dispersive infrared (NDIR) sensor is individually calibrated for operation from 0 to 50°C over the range of specified concentrations. Results are stored within each system to provide accurate analysis. It’s designed to meet or exceed the EN 14624 standard for performance of mobile leak detectors and room controllers of halogenated refrigerants.

You can also benefit from greater design flexibility thanks to the 6552’s small size and low power consumption. And it’s operating system simplifies system integration.

  • Measure up to five gases
  • Apply to any or combination of refrigeration systems that employ FreonsTM and carbon dioxide
  • Streamline system integration via a comprehensive operation system
  • Fit into various enclosures thanks to the 6522’s compact design
  • Nearly eliminate zero drift after the warm-up period

  • Compact size
  • User-defined TTL outputs for interfacing to auxiliary devices or alarms
  • Rapid warm-up
  • Pulsed-infrared source
  • Ability to measure up to five gases
  • Low power consumption

Performance

  • Response Time: Specified sample flow rate of 1 liter per minute through the 6552 sample cell
  • Data Refresh Rate: 1 second
  • Warm-Up Time: 30 min fully stable, 3 min for reduced accuracy unless zeroed prior to taking measurement
  • Compliance: Designed to meet or exceed EN 14624
  • Host Communication: USB or RS232C asynchronous serial; 19,200 bps or 9600 bps (default is 19,200)

Electrical Specifications

  • Input Power: 12 Volts DC nominal (+ 9 to + 16
  • Power Consumption: 1.8 Watts average at 12 VDC

Physical Characteristics

  • LxWxH: 19.18 x 7.37 x 5.03 cm (7.55 x (2.90 X 1.98 in)
  • Weight:: 0.3 kg (0.8 lb)

Environmental Specifications:

  • Operating Temperature Range: 0 to 70°C (32° to 158°F), accuracy not specified : >50°C
  • Operating Humidity: To 95% RH (non-condensing)
  • Operating Altitude: -300 to 3000 m (-1000 to 10,000 ft)

Customer reviews

No reviews were found for Advanced Energy - Model ANDROS 6552 - OEM Gas Analyzer. Be the first to review!