Advanced Energy Industries, Inc.
  1. Companies
  2. Advanced Energy Industries, Inc.
  3. Products
  4. Advanced Energy - Model Impac IN 140/5 ...

Advanced EnergyModel Impac IN 140/5 Series -Fast, Digital Pyrometer for Measurement of Glass Surfaces, 250 and 2500°C

SHARE

Advanced Energy’s digital Impac IN 140/5 series pyrometers are ideal for non-contact measurement of glass and quartz-glass surfaces. Short response times. Small spot sizes. Focusable optics. Digital display. Test current output.

Most popular related searches

The Impac IN 140/5 digital pyrometers are ideal for non-contact temperature measurement of glass and quartz-glass surfaces between 250 and 2500°C.

The IN 140/5 series showcases very short response time and extremely small spot sizes. It has two model variations. The IN 140/5-L is equipped with a better field-of-view (optics 2-NL) and the high-speed IN 140/5-H has a short response time of only 10 ms.

  • Accurately measure fast-moving, small objects
  • Exactly align with laser targeting light or thru-lens finder
  • Use focusable optics to match application requirements
  • View temperature measurements via the built-in display or PC
  • Adjust pyrometer parameters on the instrument or PC using InfraWin Software
  • Thru-lens view finder or laser targeting light
  • Built-in digital display
  • Interface RS232 / RS485 switchable
  • Very short response time, minimum 10 ms
  • Extremely small spot sizes, minimum 0.9 mm

Temperature Ranges

  • IN 140/5
    • 250 to 1400°C (482 to 2552°F) (MB 14)
    • 300 to 1600°C (572 to 2912°F) (MB 16)
    • 450 to 1500°C (842 to 2732°F) (MB 15)
    • 500 to 2500°C (932 to 4532°F) (MB 25)
    • Any range adjustable within the temperature range, minimum span 51°C

Measurement

  • Spectral Range: 5.14 pm (narrow band, for glass surfaces)
  • Measurement Uncertainty: See data sheet
  • Repeatability: 0.3% of measured value in °C
  • E = 1lt90 = 1s,Tamb.=23fflC
  • Exposure Time tgo: IN 140/5 and IN 140/5-L: 40 ms, adjustable up to 0.05 s, 0.25 s, 1 s, 3 s, 10 s
  • IN 140/5-H: 10 ms, adjustable up to 0.05 s, 0.25 s, 1 s, 3 s, 10 s
  • Emissivity t: 10 to 120% adjustable in the instrument or via interface in steps of 0.1%

Optical

  • Sighting: Through lens view finder or laser target light (maximum power level < 1 mW, I = 630 to 680 nm, CDRH class II)
  • Optics: IN 140/5 and IN 140/5-H: Three focusable optics IN 140/5-L: One focusable and One fixed optics
  • Distance Ratio (FOV): IN 140/5 and IN 140/5-H: minimum 150:1 (1 mm) IN 140/5-L: 180:1 (0.9 mm)

Interface and Communication:

  • Display: Illuminated LC display for temperature indication or parameter settings
  • Parameters: Emissivity, exposure time, analog output, temperature subrange, settings of the maximum value storage, address, baud
  • rate, internal temperature of the pyrometer
  • Analog Output: 0 to 20 mA or 4 to 20 mA (linear), switchable
  • Test current 10 mA or 12 mA by pressing test key
  • Digital Interface: RS232 or RS485 addressable (half duplex), switchable
  • Baud Rate: 1200 up to 115200 Bd

Environmental

  • Protection Class: IP65(DIN40 050)
  • Ambient Temperature: 0 to 70°C (32 to 158°F) at housing
  • Storage Temperature: -20to80°C(-4to176°F)
  • Weight: -0.550 kg (-1.21 lb)
  • CE Label: According to EU directives about electromagnetical immunity

Electrical

  • Power Supply: 24 VAC/DC (14 to 30 VAC/DC) (AC: 48 to 62 Hz)
  • Power Consumption: Maximum 1.2 W
  • Load: 0 to 500 0
  • Isolation: Power supply, analog output, and digital interface are galvanically isolated from each other

Typical Applications

  • Flat Glass Fabrication
  • Bulb Production
  • Car Glass Assembly
  • Glass Bending