Advanced Energy's Mikron M315X series are two-piece blackbody calibration sources with a large surface area. Temperature range: Ambient +5 to 400°C (+9 to 752°F). Manufactured and tested to meet rigid quality control standards. High emissivity and uniformity.

The Mikron® M315X series blackbody calibration sources satisfy the exacting parameters of infrared focal plane array detectors, thermal imaging, and FLIR systems testing in projection scene and field applications. 

The M315X series calibration sources are resistively heated by precision heating elements to provide uniform temperature distributions. The two-piece system is comprised of controller module and a separate enclosure for the emitter source. Two connecting cables allow 3.6 m (12 ft) of separation between the blackbody source and the controller.

  • Satisfy the exacting parameters of infrared focal plane array detectors, thermal imaging, and FLIR systems testing in projection scene and field application
  • Achieve high accuracy and resolution
  • Ensure uniform temperature distributions
  • Control the emitter source temperature with a precision digital PID controller
  • Guarantee the quality of your calibration

  • Excellent general purpose calibration
  • Large aperture sizes
  • Resistively heated by precision heating elements
  • RS232 (standard) or RS485 (option) serial communication output
  • High emissivity and uniformity

Measurement Specifications

  • Temperature Range: Ambient +5 to 400°C (+9 to 752T)
  • Temperature Uncertainty:
    • ±1°C at T<100°C to ±1.3°C at T = 400°C at 8 to 14 urn
    • Accuracy calibration performed radiometrics I ly. the uncertainty of emissivity and transfer standard are already included
  • Temperature Resolution: o.orc
  • Stability:
    • ±0.1°C per 8-hour period if in still air environment
    • Provided stable AC mains voltage and minimum air flow across the exit port or emitter plate
  • Source Non-Uniformity: See datasheet
  • Heated Emitter Shape: Flat plate
  • Exit Port Diameter:
    • M315X4: 101x101 mm (4x4 in)
    • M315X6: 152 x 152 mm (6 x 6 in)
  • Emissivity:
    • 1.00 effective at 8 to 14 um
    • Spectral emissivity graph is provided in the instruction manual
  • Standard Calibration Method: Radiometric
  • Temperature Sensor: Precision platinum RTD 1/3 DIN
  • Warm-Up Time: -30 minutes ambient to 300°C
  • Slew Rate to 1°C Stability:
    • M315X4
      • -15° per minute T < 300°C
      • ~5°per minute T > 300°C
    • M315X6
      • -10° per minute T < 350°C
      • -5° per minute T > 350°C
  • Slew Rate to 0.1°C Stability:
  • M315X4 -12 to 30 minutes between set points
  • M315X6 ~30 to 40 minutes between set points Longest near end of temperature range

Communication / Interface

  • Remote Set Point: RS232 (standard)
  • RS485 (optional)
  • Method of Control: Digital PID Controller

Environmental Specifications:

  • Operating Ambient Temperature: 10to40°C(50to104°F)
  • Cooling: Fan cooled, air inlet on rear panel
  • Operating Humidity: < 90% non-condensing
  • HxWxD: M315X4: 269 x 285 x 267 mm (10.6 x 11.22 x 10.5 in) M315X6: 417.6 x 406.4 x 371.4 mm (16.44 x 16 x 14.62 in) Controller: 195 x 432 x 576 mm (7.67 x 17 x 22.66 in)
  • Weight: M315X4:7.3kg(~16lb) M315X6: 10.4 kg (-23 lb) Controller: 9 kg (-20 lb)
  • CE Certified: Yes

Electrical

  • Power Requirements:
    • M315X4: 115 VAC at 50 and 60 Hz, 500 VA
    • M315X6: 115 VAC at 50 and 60 Hz, 2000 VA