- Home
- Companies
- Advanced Energy Industries, Inc.
- Products
- Advanced Energy - Model Mikron M305 - ...
Advanced Energy - Model Mikron M305 -Compact, General Purpose Blackbody Calibration Source, 100 to 1000°C
Advanced Energy’s Mikron M305 blackbody calibration source combines portability with a wide temperature range, high emissivity, and high accuracy. Easily moved when needed — weighs only 25 kg (55 lb). High accuracy ±0.2% of reading ±1°C. High emissivity. Manufactured and tested to meet rigid quality control standards.
The Mikron® M305 blackbody calibration source is a smart choice when portability is required and / or installation space is limited. Its spherical, metal cavity design yields an emissivity of 0.995 over the temperature range of 100 to 1000°C (212 to 1832°F) with an aperture of 25 mm (1 in). An integrally mounted PID controller adds to the versatility of this calibrator.
Custom designs and variations are available upon request.
- Easily move, weighs only 25 kg (55 lb)
- Meet rigid quality control standards
- Achieve ideal emissivity and accuracy
- Rely on fast heat-up times and temperature stability
- Integrally mounted PID controller
- Furnished with certificate of calibration traceable to NIST
- RS232 (standard) or RS485 (optional) serial communications port
- High emissivity 1.0 effective at 0.7 to 1.8 µm T ≥ 230°C, 1.0 at 8 to 14 µm T < 230°C
Measurement Specifications
- Temperature Range: 100to1000°C(210to1832°F)
- Temperature Uncertainty:
- ±0.2% of reading ±1°C
- Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included
- Temperature Resolution: ore
- Stability:
- 1°C per 8-hour period
- Provided stable AC mains voltage and minimum air flo\v across the exit port or emitter plate
- Source Non-Uniformity:
- ±0.2°C typical at T < 230°C
- ±1°C typical atT>230°C
- Heated Cavity Shape: Spherical
- Exit Port Diameter: 25.4 mm (1.0 in)
- Emissivity:
- 0.995 ±0.0005 (calculated from cavity shape)
- Effective emissivity: 1.00 at 8 to 14 urn T < 230°C, 1.00 at 0.7 to 1.8 urn T > 230°C
- Standard Calibration Method: Radiometric
- Temperature Sensor: Thermocouple
- Warmup Time: 60 minutes from ambient to 700°C
- Slew Rate to 1°C Stability:
- ~11oC/minforTArnb.<200°C
- ~20°C/min for TAmb. < 800°C
- ~10°C/minforT>900°C
- Slew Rate to 0.1°C Stability: 1 hour between set points
Communication / Interface
- Remote Set Point: Via serial port
- Method of Control: Digital, self-tuning PID controllera
Environmental Specifications
- Operating Ambient Temperature: 0 to 44°C (32 to 110T)
- Cooling: Fan-cooled, air inlet on rear panel
- Operating Humidity: 90% RH max, non-condensing
- Dimensions: H x W x D 270 x 430 x 370 mm (10.6 x 16.9 x 14.6 in)
- Weight: 25 kg (55 lb)
- CE Certified: Yes
Electrical
- Power Requirements: 115 VAC at 50 and 60 Hz, or 230 VAC at 50 and 60 Hz, 1000 W