LumaSense Technologies, Inc., An Advanced Energy Company

- Model Mikron M315-HT - Portable, Two-Piece Low Temperature Blackbody Calibration Source

From

0

Portable, Two-Piece Low Temperature Blackbody Calibration Source with a Large Surface Area, Ambient +5 to 450°C. Advanced Energy's Mikron Calibration Source M315-HT series are infrared radiators used for calibrating and verifying the output signals of pyrometers, thermal imaging systems, heat flux measurement systems, or spectrographic analysis systems. Temperature range: Ambient +5 to 450°C (+9 to 842°F). Manufactured and tested to meet rigid quality control standards. Excellent general purpose calibration.

The Mikron® Calibration Source M315-HT is a two-piece portable blackbody calibration source with a digital indicating temperature controller that can be set at any temperature between ambient 5 to 450°C (9 to 842°F). Once set, the source temperature is controlled to within ±0.3°C by an internal RTD sensor. The unit has an emissivity of 1.00 at 8 to 14 µm.

The M315-HT calibration source and the digital indicating temperature controller are contained in separate housings to allow greater flexibility in positioning the instrument within an installation (e.g. long path calibration).

  • Ensure fast heat-up times and high temperature stability
  • Experience greater flexibility in positioning the instrument within an installation
  • Achieve high accuracy and repeatability
  • Guarantee the quality of your calibration source
  • Operate with high precision and resolution

  • High effective emissivity 1.00 at 8 to 14 µm
  • Large aperture (3 in diameter)
  • Excellent stability ±0.3°C per 8-hour period
  • Furnished with certificate of calibration traceable to NIST
  • RS232 serial communication included

Measurement Specifications

  • Temperature Range: Ambient +5 to 450°C (+9 to 842°F)
  • Temperature Uncertainty: 0.25% of reading ±1°C
  • Accuracy calibration performed radio metrically, the uncertainty of emissivity and transfer standard are already included
  • Temperature Resolution: 0.1 °C
  • Stability: 0.3°C per 8-hour period Provided stable AC mains voltage and minimum air flow across the exit port or emitter plate
  • Source Non-Uniformity: Approximate ±1°C at 250°C, ±2°C at 400°C
  • Heated Emitter Shape: Flat Plate
  • Exit Port Diameter: 76 mm (3.00 in)
  • Emissivity: 1.00 effective emissivity at 8 to 14 pm
  • Standard Calibration Method: Radiometric at 8 to 14 pm
  • Temperature Sensor: Precision platinum RTD
  • Warm-Up Time: < 30 minutes from ambient to 400°C
  • Slew Rate tore Stability: Approximately 6 minutes for +50°C change
  • Slew Rate to 0.1 °C Stability: Approximately 6 minutes for +50°C change

Communication / Interface

  • Remote Set Point: Via RS232
  • Method of Control: Digital PID controller

Environmental Specifications:

  • Operating Ambient Temperature: 0to44°C(30to110°F)
  • Cooling: Fan cooled, air inlet on rear pane
  • Operating Humidity: 90% RH max, non-condensing
  • HxWxD: Blackbody Source: 207.3 x 280.4 x 266 mm (8.2 x 11 x 10.5 in)
  • Controller: 207.3 x 280.4 x 266 mm (8.2 x 11 x 10.5 in)
  • Weight: Blackbody: 4.9 kg (10.8 lb)
  • Controller: 3.2 kg (7 lb)
  • CE Certified: Yes

Electrical

  • Power Requirements: 115 VAC ±10% 50 and 60 Hz 600 W max (230 V optional)

Customer reviews

No reviews were found for Advanced Energy - Model Mikron M315-HT - Portable, Two-Piece Low Temperature Blackbody Calibration Source. Be the first to review!