- Home
- Companies
- Advanced Energy Industries, Inc.
- Products
- Advanced Energy - Model Mikron M316 - ...
Advanced Energy - Model Mikron M316 -Ultra-Portable, Low Temperature Blackbody Calibration Sources, Ambient +5 to 300°C
Advanced Energy`s M316 series is available only in the U.S. The blackbodies are two-piece, portable units ideal for field calibration or verification of infrared temperature sensors. Temperature range: Ambient +5 to 300°C (+9 to 572°F). Manufactured and tested to meet rigid quality control standards. Excellent general purpose calibration.
The compact and hand-held Mikron® M316 calibration source weighs only 0.9 kg (2.0 lb). It covers a temperature range from ambient 5 to 300°C (572°F), with an accuracy of ±0.5% of reading ±1°C. It connects to the indicating controller via a flexible cable. Due to its small size, you can easily read the sensor in hard to reach locations without altering its position.
- Ensure fast heat-up times and high-temperature stability
- Gain maximum portability and flexibility
- Achieve high accuracy and resolution
- Guarantee fast heat-up times
- Easily carry
- High effective emissivity 1.00 at 8 to 14 µm
- Large aperture (2.25 in diameter)
- Excellent stability ±0.25°C
- Fast slew rate (heat-up time)
- Furnished with a certificate of calibration traceable to NIST
Measurement Specifications
Temperature Range
Ambient +5 to 300°C (+9 to 572°F)
Temperature Uncertainty
±0.5% of reading ±1°C
Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included
Temperature Resolution
0.1°C
Stability
0.5°C per 8-hour period
Two provided stable AC mains voltage and minimum air flow across the exit port or emitter plate
Source Non-Uniformity
~ ±2°C at 250°C
Heated Emitter Shape:
Flat Plate
Exit Port Diameter
57 mm (2.25 in)
Emissivity
1.00 effective emissivity at 8 to 14 μm
Standard Calibration Method
Radiometric
Temperature Sensor
Precision platinum RTD
Warm-Up Time
~ 10 minutes from room temperature to 200°C
Typical - may vary from unit to unit
Slew Rate to 1°C Stability
- ~ 18° per minute T < 250°C
- ~ 14° per minute T > 250°C
Typical — may vary from unit to unit
Slew Rate to 0.1°C Stability
Approximately 12 minutes between Δ 50°C set points
Communication / Interface
Method of Control
Digital PID Controller
Environmental Specifications
Operating Ambient Temperature
0 to 50°C (32 to 122°F)
Cooling
Self-convection
Operating Humidity
< 90% non-condensing
H x W x D
- Blackbody source: 203 x 89 x 98 mm (8 x 3.5 x 3.86 in)
- Controller: 102 x 178 x 127 mm (4 x 7 x 5 in)
Method of Mounting
- Blackbody source: handle grip and 1/4-20 tapped hole for tripod mount
- Controller: bench mount or handle for field use
Weight
- Blackbody source: 0.9 kg (2.0 lb)
- Controller: 0.70 kg (1.5 lb)
CE Certified
No
Electrical
Power Requirements
115 VAC at 50 and 60 Hz 300 W maximum 1