airmoModel S -Automatic Isothermal Gas Chromatograph Analyzer

SHARE

An autoGC-trap-FPD for low sulfur quantification. airmo S is a fully automatic isothermal gas Chromatograph for the ana­lysis & monitoring of sulfur compounds with Sulfur Specific Detector. This solution is specific for application where low sensitivity on sulfurs must be reached. This auto GC uses a valve with 1 sample trap, and metallic capillary columns.

Most popular related searches

The main specifications are :

  • Continuous, uninterrupted sampling
  • Sample is injected into column oven for separation
  • Controlled oven selects optimum temperature based on applica­tion
  • Detection achieved by a sulfur specific photometric double flame detector (FPD)
  • Automatic data transfer through:
    • Modbus or MSG1 communication protocol
    • Analog 4-20mA or 0-10V output
  • Hydrogen Generator
  • Zero Air Generator
  • Multiple stream analysis with Multiplexer
  • Dilution system
  • Heated lines
  • N2 Generator
  • Dual mode chromaS/airmoS

Compounds analysed:

  • Specific sulfur compounds analysis:
  • Me-SH, Et-SH, DMS, DES, DMDS, CS2 …
  • Options: n-propyl-SH, 2-butyl-SH, Isopropyl-SH, Terbutyl-SH
  • Only with loop in option: H2S, SO2, COS

Detection limits with trap:

  • 0,1 ppb double Sulfur Compounds like DMDS, CS2 etc...
  • 0,2 ppb single Sulfur Compounds (MM, DMS etc..)
Detection limits with loop in option:
with loop = 250 μl and amplification 2
  • 14 ppb for H2S/SO2/ MM/EM/DMS/DES/COS
  • 8 ppb for DMDS & CS2

Detection range (selectable ppm / ppb)

  • 0,2 to 10 ppb (amplification 2)
  • 1 to 40 ppb (amplification 1)
Detection range with loop in option (selectable ppm / ppb)
ppm or ppb : amplification 1&2
  • 0,01 to 10 mg/m3 or 0 to 600 mg/m3
  • % / ppm / ppb level depends on compounds for ex H2S = 0,01 to 10 mg/m3 = 7 ppb to 7 ppm

Relative Standard Deviation:

  • RSD < 5% on concentration over 24hrs
  • RSD < 1% on Retention Time over 24hrs

Supervisor:

  • Full result storage (data and chromatogram)
  • Embedded computer Windows® based with LCD display
  • 128 GB of Hardware storage on SSD memory
  • 4 USB Connecting Port
  • Two RS-232 ports
  • Display: 10’’ TFT Color LCD

Cycle Times:

  • 10 min in airmoS mode or chromaS mode
  • 30 min in dual mode (option): 10 min chromaS + 20 min airmoS

Gas Supply:

  • Carrier Gas - N2 recommanded / Air possible 3 bar (5ml/min) 1/8’’
  • Sample Inlet -1 Bar or ambient pressure (airmopump)
  • Sampling dryer: 150-200 ml/min of dry air
  • Detector
    • H2 at 2 Bar (70 ml/min) 1/8’’
    • Air at 3 bar (80 ml/min) 1/8’’
Operation Temperature:
  • Room with air conditioning: 10 to 25°c

Power Supply:

  • Main: (230V / 50Hz) or (115V / 60Hz)

Electrical consumption:

  • 250W (+50W for computer)

Dimensions and Weight:

  • Rack: 482 mm (19’’)
  • Height: 177,8 mm (4U) or 222 mmm (5U)
  • Depth: 660mm