3S Analyzers S.r.l.

3S-AnalyzersModel CL -Aluminum Colorimetric Online Analyzers

SHARE

The 3S colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer`s glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).

Most popular related searches
  • Drinking water
  • Industrial waste water
  • Municipal waste water
  • Surface water
  • Low reagents consumption and analytical reliability thanks to our custom designed miniature peristaltic pumps
  • Dual compartment design to guarantee a complete separation between the electronics and the hydraulic section
  • Color touchscreen to show the last measured value clearly, and to set all the options and settings via an user friendly interface
  • Datalogger function with data history in graphical form and USB download
  • Automatic calibration, validation and cleaning to reduce down time and operator intervention, their frequency can be freely set by the user
  • Fully integrable in industrial automation via the analog outputs (2 x 4-20 mA channels), digital output (2 relays + ModbusRTU) and digital input (voltage free contact)

In a pH 6.2 to 6.4 buffered solution pyrocatechol violet and Al(III) form a blue colored compound. The absorbance intensity is proportional to the aluminum concentration in the sample and is measured at 572 nm.

Reagents

Standard solutions

  • Range (Ø 26 mm) 150 ppb
  • Range (Ø 16 mm) 250 ppb
  • Range (with dilution) 20 ppm
  • Wavelength 572 nm
  • Number of reagents 2
  • Min. analysis time 8 minutes
  • Model CL
  • Sampling Mode Batch, with freely settable frequency.
  • Sample Pressure: Atmospheric
  • Temperature: 5 ‐ 45°C (41 ‐ 113°F)
  • Flow Rate: 80 to 500 mL/min
  • Connection: 6 mm (¼-in.)
  • Drain Pressure-free drain
  • Connection: 12 mm (½-in.)
  • Input Streams 2
  • Cabinet Material Epoxy-coated stainless steel AISI 304
  • Dimensions (H x W x D) 604 x 380 x 242 mm (23.6 x 14.8 x 9.4 in)
  • Weight 20 kg (44 lbs)
  • Power Supply Input Voltage:115 - 240 VAC 50/60 Hz or 24 VDC (optional)
  • Power consumption: max. 80 W
  • Output signals 2 x 4-20 mA analog outputs
  • Modbus RTU via RS485
  • Alarms 2 SPDT programmable voltage-free relays
  • Digital Input Online,Start Extra,Skip Idle,Emergency Stop
  • Ambient Temperature 5 - 45 °C (41 - 113 °F)
  • Ambient Humidity 10 to 90 % RH (indoor use only, outdoor installation possible with protective cabinet or shelter not included)
  • Protection Grade IP54