ChemScan - Model mini LowAm -Ammonia Analyzer
FromIn-Situ Process
The single parameter in-line analyzer family utilizes years of ChemScan® experience and proven technology to provide reliable and accurate analysis of water and waste water. This device has been designed from the ground up to reduce maintenance requirements, includes large ID sample tubing to minimize plugging and only needs quarterly reagent refills.
Most popular related searches
automatic analysis
sample handling
sampling tube
ammonia analyzer
effluent water
field sampling
total suspended particulate solids
effluent waste
water analysis
- High Reliability
- Low Capital Cost
- High Accuracy
- Low Operating Cost
- Waste Water Effluent
- Automatic Analysis Utilizing ChemScan`s Proprietary Method
- Low Maintenance
- Proven Sample Handling with Large I.D. Flow Paths
- Simple Field Adjustable Calibration
- Sample Blank to Eliminate Backgound Interference
- Automatic Cleaning
- Automatic Analysis
- Continuous Output
- Multiple Data Communication Interface Options
- Ammonia
Range (as N):
0.01 - 10.0 mg/L (Method 1066)
Accuracy:
2% of value or 2x detection limit (whichever greater)
Cycle Time:
15 minutes to 9999 minutes (field programmable)
Environment:
5 - 50 degrees C (method dependent)
Power:
100 - 240 VAC, 50 W
Enclosure:
NEMA 4x
Safety Approval:
CSA-US
Sample Requirements:
0.5 - 1 Liter/analysis, pressure 2 to 10 psi,
<150 mg/L TSS, <60 NTU
Maintenance:
Reagent replacement every 3 months, pump kit yearly
Relay Contacts:
1 SPDT Concentration, 1 SPDT Programmable
Serial Interface:
Serial, RS-232 Maintenance Port
Analog Output:
Isolated 4-20 mA