Applications: Climate and environmental research / Trace gas analysis. Online monitoring of HCHO emission / Emission limit control. Process control / Quality control for the chemical industry. Chamber testing of wood-, plastic-, fabric-based products using up to 12 chambers simultaneously. Measurement of indoor air quality. Replacement for the wet chemistry in the acetyl-acetone method.
Most popular related searches
indoor air quality
indoor air quality control
emissions test chamber
environmental air monitoring
trace gas analysis
air quality control
indoor air
indoor air quality monitoring
emissions control
liquid sampling
- HCHO detection technique: Fluorimetric, using Hantzsch (acetyl-acetone) reaction
- Linear detection range: 0.1ppb to 3000ppb (gaseous), 150ng/liter - 5mg/liter (liquid)
- Detection limit:
- Sample gas temperature: >0°C to +120°C
- Calibration and zeroing: Automatic zeroing and semi-automatic calibration using liquid standards or automatic using internal gas generator (optional)
- Weight and dimension: 20kg, fit for 19" rack (whd: 45cm×15cm×56cm)
- Power requirements: 110VAC / 220VAC, 110W, 24VDC on request
- Continuous online monitoring of formaldehyde emissions with unique sensitivity of 100ppt
- Analysis of gaseous and liquid samples with only one instrument
- Designed for climate research, environmental air monitoring and indoor air quality control
- Ideal for emission control of wood-, plastic-, and fabric based products
- Can be employed together with up to 16 emission test chambers