- Home
- Companies
- Microlab Instruments Private Limited
- Products
- Microlab - Model ARX 199/ARX 199i - ...
Microlab - Model ARX 199/ARX 199i -Biochemistry Analyzer
User friendly software, Editing of patient ID, Name (12 characters), Factor, Print out of laboratory name & Doctor`s name. For Kinetic test : On line Graph & OD. Graph editing facility. Estimation of result. Universal input voltage 90 to 265 V/50 - 60 Hz. Storage of 2000 test results in the memory with patient ID. PS2 Keyboard connection optional. Lamp Saver mode.
METHODS OF ASSAY
- ABS
- EP with Factor
- KIN
- Fixed time
- Differential
- Multipoint upto 10 standards
OPTICAL SYSTEM
Principle: Absorption spectrometry
Range of measurement: From -0.1 to 3.0A
Normal Spectral range: Automatic via filter wheel
Filters: 340, 405, 505, 546, 578, 630, 670 nm. 1 extra position
Bandwidth: 10nm (±2 nm)
Reading error: 2% at 2.5 A
Light source: Halogen lamp
Zero: Automatic
Settings: Monochromatic & Bichromatic readings.
Detector: Schottkey type photodiode - Silicon
THERMOSTAT SYSTEM
Peltier regulated: 25°, 30° & 37°C
Programmed Accuracy: 0.2°C
CUVETTES - DUAL MODE
Flow cell of: 18 µl
Removable cuvettes: Macro, Semi-micro & Micro
PRINTER
Type: Fast Thermal Graphic
Printing speed: 50 mm/sec
Dots per line: 384 dots/line
Paper: Thermo sensitive, 57 mm wide
SCREEN AND KEYBOARD
LCD blue LED back lit: 240 x 64 pixels.
Keyboard: Tactile Soft-touch
ASPIRATION SYSTEM
Continuous flow system with Bi-directional peristaltic pump
INCUBATOR
37°C for 9 Position Test Tubes & 9 Position Cuvettes
PROGRAMMING
Table of tests: up to 200
Table of units: up to 9 units
Quality control storage of the last 30 results
Memory: 128 KB EEPROM
QUALITY CONTROL
2 Controls per test - Normal & Abnormal
POWER / DIMENSIONS
Voltage: 110-220 V
Frequency: 50 - 60 Hz
Maximum power: 80W
Temperature: 10° to 40° C
Maximum relative humidity: 80%
Dimensions: 40 cm x 32 cm x 16