CAI - Model 700 LX HCLD -Heated Chemiluminescence Analyzer
The California Analytical 700 LX Series HCLD Analyzer utilizes the principle of chemiluminescence for analyzing the NO or NOx concentration within a gaseous sample. In the NO mode, the method is based upon the chemiluminescent reaction between ozone and nitric oxide (NO) yielding nitrogen dioxide (NO2) and oxygen. This reaction produces light which has intensity proportional to the mass flow rate of NO2 into the reaction chamber
The light is measured by means of a photodiode and associated amplification electronics. In the NOx mode, NO plus NO21S determined as above, however, the sample is first routed through the Internal N02 to NO converter which converts the NO2 in the sample to NO. The resultant reaction is directly proportional to the total concentration of NOx.
- Measures From 3 ppm up to 3,000 ppm Full Scale
- Three Measurement Modes—NO, NOx and NO/N02/NOx
- Heated Oven Allows Hot/Wet Sampling
- Auto Calibration and Ranging
- Fast Response Time
- Robust and Rugged Linux Based Operating System
- Electronic Sample and Ozone Flow Control
- Does Not Require Vacuum Pump
- Compreshesive Diagnostics
- Standard Outputs: Voltage, Current,
- AK Protocol, RS-232 AK Protocol, TCP/IP
- MODBUS
- CE Mark and ETL Listed—Conforms to UL STD 61010-1, Certified to CAN/CSA C22.2 STD No. 610610.1
- 1065 and ECE 49-06 Compliant
- Combustion Efficiency
- Fuel Cell Analysis
- Turbine/Generator Feedback Control
- Ammonia Slip
- Process Chemical Gas Analysis
- Personnel Safety
- Pharmaceutical Processes
- Vehicle Emissions