Chlorophyll Online Analyzer

SHARE

The principle of PCH-800 Chlorophyll Sensor is using the characteristics of chlorophyll A who has absorption peaks and emission peaks in the spectrum.The absorption peaks emit monochromatic light into the water,chlorophyll A in the water absorbs the energy of monochromatic light,releasing monochromatic light of emission peak of another wavelength.The light intensity emitted by cyanobacteria is proportional to the content of chlorophyll A in water.

Most popular related searches
  • Based on the Fluorescent measuring target parameter of pigment,can be identified before affected by potential water bloom.
  • Without extraction or other treatment, rapid detection to avoid the impact of long shelving the water sample.
  • Digital sensor,high anti-jamming capacity and far transmission distance.
  • Standard digital signal output,can achieve integration and networking with other equipment without controller.
  • Plug-and-play sensors,quick and easy installation.
  • Chlorophyll online monitoring of water plant inlet,drinking water source and aquaculture and etc.
  • Chlorophyll online monitoring of different water bodies,such as surface water,landscape water and seawater and etc.

Measurement range: 0-500 ug/L

Measurement Accuracy: ±5% of the signal level corresponding value of 1ppb Rhodamine WT Dye

Pressure range: ≤0.4Mpa

Calibration: Deviation value calibration,slope Calibration

Requirements: Suggest a multipoint monitoring for the distribution of Blue-Green Algaein water is very uneven.Water turbidity is below 50NTU.

Main material: Body:SUS316L(fresh water),Titanium alloy(marine);Cover:PVC ;  Cable:PVC

Power supply: AC:100-240VAC(50/60HZ), DC:90-36VDC

Output: 3-way 4-20mA

Relay: Setting three-way relays,programmed response parameter and response value.

Communication protocol: MODBUS  RS485

Storage temperature: -15- 65?

Operation temperature: 0- 45?

Dimension: Sensor:Dia37mm*L 220mm, Sensor:145*125*162mm(L*W*H)

Weight: Sensor:0.8KG;Transmitter:1.35KG

Protective rate: Sensor:IP68/NEMA6P;Transmitter:IP65/ NEMA4X

Cable length: Standard:10m,the maximum may be extended to 100m