Qubit SystemsModel Q-S153 -CO2 Analyzer

SHARE

Q-S153 CO2 Analyzer is a single channel non-dispersive infrared CO2 analyzer which measures CO2 in 0 to 5% and 0 to 10% range with 0.01% resolution. Q-S153 replaces our S153 CO2 Analyzer. The same dependable technology has been improved, made more rugged and modular for easy fit in our  Q-Box Packages. Q-S153 CO2 analyzer is ideal for CO2 exchange measurements in larger or more active animals and in situations where high CO2 fluxes may occur, such as fermentation processes, human respirometry. Q-S153 may be used in a flow-through gas exchange configuration for instantaneous and continuous measurements of CO2 exchange or as part of Qubit Systems’ carbon dioxide control system for regulating pCO2 in growth cabinets and rooms.

Most popular related searches

Packages that include Q-S153 CO2 Analyzer:

  • Switchable ranges of 0 – 5% and 0 – 10% CO2
  • 0.01% CO2 resolution on digital display
  • Non-dispersive infrared technology
  • Modulated infrared light source = no moving parts
  • 0 – 5 V analog output at both range settings
  • Optional battery pack for field use
  • Compact and portable

  • Respirometry in mammals
  • Exhaled air analysis in humans
  • Compost monitoring
  • Monitoring of anaerobic metabolism in yeast colonies

  • Operating principle: Non-dispersive infrared
  • Gas sampling mode: Flowing gas stream, sealed chamber
  • Maximum gas flow rate: 650 mL/min
  • Measurement range (LCD display): 0 – 10%
  • Analog output, low sensitivity: 0 – 10%
  • Analog output, high sensitivity: 0 – 5%
  • Accuracy: ± 0.3% of full scale
  • Repeatability: Better than ±1%
  • Maximum drift (per year): ±0.5%
  • Response time (@ 250 mL/min; to 95% of final value): ca. 20 sec
  • Warm up time 5 min
  • Output (linear) for Low Sensitivity setting: 0 – 5 VDC for 0 –10%
  • Output (linear) for High Sensitivity setting: 0 – 5 VDC for 0 – 5%
  • Calibration adjustments: Zero and Span
  • Operating temperature range: 0 to 50C
  • Storage temperature range: -40 to 70C
  • Operating pressure range: ±1.5% local mean pressure
  • Humidity range: 5 to 95% RH, non-condensing (recommend drying gas stream)
  • Pressure dependence: +0.19% reading per mm Hg
  • Power requirements: 12 VDC via 120 VAC/60 Hz adapter
  • Current requirements: 175 mA average, 450 mA peak
  • Dimensions (cm): (H x W x D: 5.5 to 9.5 x 9.5 x 17)
  • Weight: 1 kg
  • Warranty: 1 year limited