GRIMM Aerosol Technik GmbH & Co. KG - a member of the Durag Group

GRIMMModel MiniWRAS 1371 -Compact Mini Wide Range Aerosol Spectrometer

SHARE

The compact Mini Wide Range Aerosol Spectrometer (MiniWRAS) is the only portable instrument on the market that allows simultaneous and precise real-time monitoring of both micron-sized and nano particles. Designed and specifically built for indoor air quality monitoring, the MiniWRAS is a fit for purpose, state-of-the-art system that combines optical and electrical particle detection in one single device. This system offers many features, such as a 10 nm to 35 µm ultra-wide particle size range, simultaneous PM1, PM2.5, and PM10 measurement, 41 high resolution particle size channels, remote data transmission and instrument control. Easy to use, transport and deploy, it can be flexibly applied in various IAQ monitoring projects. 

Most popular related searches

The MiniWRAS is the optimal solution for monitoring indoor PM values taking the nanoparticles into account and revealing the particle size distribution in the small nanometer range. It´s a perfect tool for indoor air quality monitoring at working places, inside vehicles, airplane cabins, trains, R&D testing, and many more.

  • ultra-wide size range from 10 nm to 35 μm
  • PM10, PM2.5, PMand particle size distribution, particle surface, and dust mass
  • high precision over 41 equidistant channels
  • no consumables
  • non-radioactive particle charger
  • versatile data aquisition and communication interfaces (Bluetooth, USB, RS-232)
  • easy to use with GRIMM software
  • optional sensor for temperature and relative humidity
  • self-test of all optical and pneumatic components for high quality standards
  • rinsing air for protecting laser and detector in optical cell
  • nanoparticle and PM monitoring (e. g. PM2.5)
  • Indoor Air Quality (IAQ) in buildings
  • IAQ in vehicles, airplane cabins, cockpits, busses, trains
  • nanoparticle source identification
  • workplace monitoring
  • R&D testing in industry
  • measured parameters : dust fractions acc. to EN 481 (inhalable, thoracic, respirable) PM10, PM2.5, PM1,
  • number concentration and size distribution
  • dust mass : 0 – 100 000 μg/m3
  • particle size range : 10 nm – 35 μm (10 – 193 nm electrical, 0.253 – 35 μm optical)
  • size channels : 41 (10 electrical and 31 optical)
  • particle number
    • 3 000 – 500 000 p/cm³ (electrical)
    • 0 – 3 000 000 p/L (optical)
  • reproducibility : >97% of total measuring range (optical)
  • detection principle optical : light scattering at single particles; detection volume aerodynamically focused, no border zone error
  • optical cell : diode laser 660 nm
  • detector : fast signal processing with 2 μs pulse length, 2 x 16 raw data channels
  • time resolution : 6 s, 31 channels (storage interval 1 min)
  • detection principle electrical : electrical mobility spectrometer with Faraday Cup Electrometer
  • detector sensitivity : 0.25 fA
  • time resolution : 60 s, 10 channels 6 s each (storage interval 1 min)
  • volume flow : 1.2 L/min, ± 3% constant due to self-regulation
  • internal rinsing air flow rate : 0.4 L/min, protects laser optics, reference air for self-test
  • operation : GRIMM MiniWRAS software (wireless or data cable)
  • interfaces : Bluetooth, USB, RS-232
  • analog input :  external sensor for temperature and relative humidity
  • power supply: in: 100 – 240 VAC, 47 – 63 Hz, out: 18 VDC, 2.5 A
  • battery Li-lon battery, 14.4 VDC, 4.8 Ah for 8 h operation : 
  • dimensions (h x w x d) : 34 x 31 x 12 cm (13.4 x 12.2 x 4.7 in)
  • weight : 7.6 kg (16.8 lbs)
  • operating conditions : +4 to +40°C (39 - 104°F), RH < 95%, non-condensing