SAXON Junkalor GmbH
  1. Companies
  2. SAXON Junkalor GmbH
  3. Products
  4. SAXON - Model NDIR8000 - Compact, ...

SAXONModel NDIR8000 -Compact, User-Configurable NDIR Measuring Bench

SHARE

The NDIR8000 is a compact, user-configurable NDIR measuring bench for special applications - like TOC. Analog outputs 4 … 20mA integrable, special version for TOC analyzers available, location-independent operation, permanent temperature and pressure compensation, reference channels for maximum precision, electrochemical cells integrable.

Most popular related searches
  • multi-channel NDIR measuring bench for up to 6 gases
  • TOC version with high resolution 100ppm CO2 channel
  • Outputs RS232 and analogue 4-20mA (Option)
  • for direct integration in your analyzers and process controls
  • CO: 5000 vppm, 10 % Vol., 40 % Vol.
  • CO2: 100 vppm, 1000 vppm, 10 % Vol., 100 % Vol.
  • C3H8: 2.000 vppm, 5.000 vppm, 2 % Vol.

..more on request.

  • Operating Mode: NDIR
  • Output
    • digital: RS232 (12Bit ADC)
    • analog: 4…20mA (Option)
  • Operating Temperature: 5…55°C:
  • Mass: 0,8 kg:
  • Operating Voltage:
    • 15VDC…24VDC
    • 100mA @ +24VDC (TOC-Version)
    • 150mA @ +24VDC (Doppelküvette)
  • Dimensions: 320 x 111 x 60 mm (Länge x Breite x Höhe)
  • Zeropint Drift: < 2% of the Range / 24h
  • Sensitivity Drift: < 2% of the Value / Woche
  • Temperature influence: < 0.1% of the Range / K
  • Linearität: < 3% of the Range
  • Air Pressure: < 0,1 % of the Range / hPa (760…1160hPa)
  • Ripple: < ± 0,2 ppm im Bereich 500 ppm CO2
  • gas connections: Schlauchkupplung DN 4/6 für DN 4/6 Schläuche
  • Flow:
    • Constant flow necessary
    • Range 20…100l/h, max. variation ± 2l/h
  • erlaubter Gasdruck: 20…200hPa
  • Warm-Up Time