MadgeTech, Inc.
  1. Companies
  2. MadgeTech, Inc.
  3. Products
  4. MadgeTech - Model UltraShock - Data ...

MadgeTechModel UltraShock -Data Logger

SHARE

The UltraShock is a temperature, humidity, pressure and tri-axial shock data logger. Includes certificate of conformance.

Most popular related searches

The new UltraShock contains powerful improvements and an all-in-one functionality, allowing it to replace all current shock models and their capabilities. New refinements also include a smaller, lighter design and LED status indicators.

Additional features include a significantly improved battery life and the use of a USB-C cable, making it easily rechargeable. Each UltraShock purchase contains an AC adapter and USB cable, eliminating the need for an additional interface cable.

The UltraShock is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

General

Memory: 4,000,000 readings (660,000 per channel, all channels in use)

Reading Rate:

  • 1024 Hz to 5 minutes for shock, selectable in software. Temperature, pressure & humidity sampled approx. every
  • 2 seconds at intervals shorter than
  • 2 seconds. Otherwise, sampled at the reading rate.

Battery Life: 90 days typical at 64Hz rate

Material: Anodized aluminum

Dimensions: 3.4 in x 1.7 in x 1.3 in

(86 mm x 43 mm x 33 mm)

Operating Environment: -20 ºC to +60 ºC, 0 %RH to 95 %RH non-condensing

Required Interface Package: USB (type C cable required); 1MBaud

Approvals:

  • CE compliant
  • EMC Directive 2014/30/EU
  • RoHS Directive 2011/65/EU
  • IP Rating 64

Temperature

Sensor: Semiconductor

Range: -20 ºC to +60 ºC

Resolution: 0.1 ºC

Accuracy: ±0.5 ºC (+0 ºC to +50 ºC)

Humidity

Sensor: Capacitive Polymer

Range: 0 %RH to 95 %RH

Resolution: 0.1 %RH

Calibrated Accuracy: ±3 %RH (±2 %RH typical at 25 ºC)

Specified Accuracy Range: 0 °C to +50 °C; 20 %RH to 80 %RH

Pressure

Sensor: Semiconductor Strain Gage

Range: 100 to 1300 mbar

Resolution: 0.05 mbar

Calibrated Accuracy: ±1.5 mbar at 25 ºC; at 750 mbar

Shock

Accelerometer Type: MEMS Semiconductor

Acceleration Range (g): ±15 g/±100 g/±300 g

Acceleration Resolution (g): 0.02 g/0.05 g/0.2 g

Calibrated Accuracy (g): ±0.3 g/±2.0 g/±6.0 g

Sampling Rate: 1000 Hz

Accelerometer Freq. Resp.:

  • >1000 Hz (15 g)
  • >500 Hz (100 g, 300 g)

Reading Rate:

  • 1024 Hz to 5 minutes for shock, selectable in software. Temperature, pressure & humidity sampled approx. every
  • 2 seconds at intervals shorter than
  • 2 seconds. Otherwise, sampled at the reading rate.