Mettler - Toledo Int. Inc
Mettler Toledo - Degassed Cation Conductivity System
The METTLER TOLEDO Thornton DCC1000 System offers a new design for conductivity measurement for power cycle chemistry monitoring. By providing conductivity measurement in compliance with ASTM D4519, this system provides assurance of water purity to maximize power production and minimize corrosion. Unambiguous measurement of trace levels of corrosion-causing contaminants is enabled with minimal operator supervision.
Most popular related searches
conductivity monitoring
sample temperature
sample conditioning
flow sensor
degassing system
integrity monitoring
corrosion monitoring
trace contaminant
trace monitoring
- Precise detection of corrosive contaminants for faster plant start-ups and simpler turbine warranty compliance
- Multi-parameter M800 transmitter with single-screen display of all measurements for easy displaying and monitoring of sample conditions
- Integrated flow sensor with automatic heater shut-off if flow stops to protect the DCC1000 system from thermal damage and maintenance
- Trend graphs for all measurements to understand plant characteristics better and plan maintenance, avoiding unplanned shutdowns
Specifications - Degassed Cation Conductivity System
- Power supply: 100-140 VAC and 200-240 VAC, 975 W typical
- AC frequency: 50 to 60 Hz
- Sample flow rate: 100-250 mL/min
- Sample temperature: <100 °C (212 °F)
- Sample pressure: 0.3-7 bar (5-100 psig)
- Cation resin: 1 L
- Ambient operating temperature: 5-50 °C (41-122 °F)
- Humidity: 10 - 90% non-condensing
- Dimensions (HxWxD): 850 x 800 x 170 mm
- (33.5 x 31.5 x 6.7”)
- Weight: 20 kg (44.1 lb)
- Rating / approvals: CE