Harvard Apparatus
  1. Companies
  2. Harvard Apparatus
  3. Products
  4. Harvard - Model 33 DDS - Dual Drive ...

HarvardModel 33 DDS -Dual Drive Syringe Pump System

SHARE

The Harvard Apparatus Pump 33 DDS (Dual Drive System) is a leap forward in syringe pump capability. The Pump 33 DDS has two independent pumping channels controlled by an intuitive touch screen interface.

Most popular related searches

This multi-purpose syringe pump employs advanced syringe mechanisms that include a tight gripping, extremely secure syringe clamp that accommodates syringe sizes 0.5 µl to 60 ml. The Pump 33 DDS offers enhanced flow performance with high accuracy and smooth flow from 1.02 pl/min to 106 ml/min.

The new Harvard Apparatus Pump 33 DDS employs a graphical user interface controlled with a large 7” LCD Color Touchscreen.

Syringe tables containing all major syringe manufacturers allow simple selection of any compatible syringe size.

Audible Alarms, Adjustable Force and Screen Lock are all features that are available with the touch of the screen.

The Harvard Apparatus Pump 33 DDS (Dual Drive System) is a leap forward in syringe pump capability.

The Pump 33 DDS has two independent pumping channels controlled by an intuitive touch screen interface. This multi-purpose syringe pump employs advanced syringe mechanisms that include a tight gripping, extremely secure syringe clamp that accommodates syringe sizes 0.5 µl to 60 ml. The Pump 33 DDS offers enhanced flow performance with high accuracy and smooth flow from 1.02 pl/min to 106 ml/min.

This new syringe pump employs a graphical user interface controlled with a large 7” LCD color touch screen for quick and easy syringe pump setup. Syringe tables containing all major syringe manufacturers allow simple selection of any compatible syringe size. Audible Alarms, Adjustable Force and Screen Lock are all features that are available with the touch of the screen.

The bright informative display run screen presents the user with all key dispensing parameters in real time.

  • Two Independently Controlled pumping channels in one instrument
  • Run two separate flows using different directions,  flow rates, volumes and syringe sizes
  • Intuitive Icon Based Graphical User Interface controlled with a 7” LCD Color Touch Screen Display
  • Accommodates Syringe Sizes 0.5 µl to 60 ml
  • Smooth flow down to 1.02 pl/min
  • High accuracy ± 0.25%
  • USB, RS-232 & TTL Connectivity
  • Accuracy± 0.25 %
  • Linear Force
  • (Max, per syringe)
  • 70 lb (31.75 kg) at 100% Force Setting up to a flow rate of 90 ml/min using up to a 60 ml syringe with a 32.573 mm inner diameter. 50 lb (22.6 kg) at 100% Force Setting for flow rates 90 ml/min to 106 ml/min using the same size syringe.
  • Syringe Size:
  • Two Independent syringe mechanisms (noted as syringe drive P1 & P2)
  • Minimum0.5 µl(0.103 mm minimum inner diameter)
  • Maximum60 ml(32.573 mm maximum inner diameter)*
  • Flow Rate:
  • Minimum1.02 pl/min (0.5 ul syringe, 0.103 mm inner diameter)
  • Maximum106 ml/min (60 ml syringe, 32.573 mm diameter )
  • Display 7" WQVGA TFT Color Display with Touch Screen
  • Modes of Operation:
  • Twin ConditionBoth syringes of the same size operate identically (flow rate, direction & volume)
  • Independent ConditionBoth syringes operate independently
  • Reciprocating ConditionContinuous flow, both syringes of the same size operate identically in opposite directions
  • Non-Volatile Memory Stores all settings
  • Pump Command ControlASCII Command Set
  • Real time ClockYes, with battery backup (battery included and required for real time clock)
  • Connectors:
  • PowerBarrel connector, (-) barrel (+) post 2 mm X 5 mm male plug
  • RS-485 IEEE-1394, 6 pos (Pump-to-pump communication only)
  • USB Type B
  • RS-2329 pin D-Sub Connector
  • (I/O) TTL15-pin D-sub connector (one for each axis)
  • Footswitch ConnectionsMini phono jack
  • Drive MotorTwo independent stepper motors
  • Motor Drive Control MCU controlled
  • Step Rate:
  • Minimum 27 sec/µstep
  • Maximum26 µsec/µstep
  • Stall DetectionYes,independent axis stall detection
  • Input Power30 V, 2.0 A
  • Power Supply Input 100 to 240 VAC, 50/60 Hz, Output 30 V 2.0 A, 50 Watts
  • Dimensions (L x D x H)11 x 15 x 8 in (28 cm x 39 cm x 21 cm )
  • Weight 21 lb (9.09 kg )
  • Operating Temperature 4°C to 40°C (40°F to 104°F)
  • Storage Temperature -10°C to 70°C (14°F to 158°F)
  • Operating Humidity 80% @ 25° C ambient temperature
  • Storage Humidity 20% to 80% RH, non-condensing
  • Classification Class I
  • Pollution Degree 1
  • Installation Category II
  • Regulatory Certifications CE, ETL (UL & CSA), CB Scheme, EU RoHS, WEEE