E-Val - Model Pro -Sensor Interface for Additional Measuring Parameters
To reach the required pressure accuracy described by the EN 285 norm, the E-Val Pro sensor interface can, among other things, be equipped with Ellab’s specially designed 0-4 Bar sensor. Thereby providing the required accuracy of +/- 10 mBar. In addition to pressure measurements, the E-Val Pro Sensor Interface is the ideal solution for expanding the number of measurable parameters available for wired thermocouple systems. TrackSense Pro Data Logger sensors can be mounted on the interface, allowing it to measure Temperature, Vacuum, Relative Humidity (EtO), Conductivity and CO2, all while being connected to the E-Val Pro.
With the ValSuite software and the Sensor Interface, the calculated parameters can be displayed on a PC and directly on the E-Val Pro display.
Using multiple sensors in conjunction with the Sensor Interface (i.e. Double, Quad or Combi sensors) is also a possibility as long as the sensor array channel(s) next to the interface are free.
Interface
- Communication Protocol: RS485
- Measuring Range: Defined by connected sensor type
- Operating Temperature (ambient): -40 to +125 °C
- Operating Pressure: 0.001 mBar to 10 Bar
- Material: Stainless Steel 316L
- Vacuum, CO2 and Relative Humidity (EtO) Sensor Requirements: Connection via 4-Channel Sensor Array
- Temperature, Pressure, Relative Humidity, Relative Humidity (EtO/Stability) and Conductivity Sensor Requirements: Connection via 4 or 12-Channel Sensor Array
- ValSuite: ValSuite 5.1 or later
- Adaptors: 1/2″ or 3/4″ thread adaptors
- Calibration: Stored in interface until the sensor is changed
- Sample Rate: 1 sec. to 24 hours
- E-Val Pro Main Firmware Requirements: 4.4.0.3 or later
- Array Firmware Requirements: 1.12 or later
Cable
- Operating Temperature: -40 to +70 °C
- Length: 5 m
- Diameter: 4 mm
- Material: PVC
- Plug Type: USB