Furnace Internal size: 9" width X 9" height X 18" depth. Furnace hot zone volume : 9" width X 9" height X 18" depth. Furnace entrance size: 9" width X 9" height. Insulation : High grade pressed zirconium Blanket ceramic Fiber and insulation brick, insulation thickness will be 250mm approx. Chamber: Chamber made of high alumina cumilite brick 90% alumina temp 1600 max. Front Loading with Suitable Opening Door Mechanism. Shell Construction: Double Walled High quality fabrication of M.S.Body. Angle frame 40/4 and M. S. sheet 14 SWG, Front Plate 8 mm Thickness Structure with Proper Stiffeners and neatly powder coated. Door Construction: Sophisticated system to avoid heat loss with proper insulation.
Most popular related searches
digital temperature controller
cement testing system
high temperature furnace
furnace
cement testing
ceramic fiber insulation
heat safety
ceramic fiber
temperature sensor
ceramic fiber blanket
- Heating Elements: Silicon carbide rood (12 numbers with even Distribution) Element is Two Side left & Right
- Type of element: Solid type
- Hot zone length: 200 mm
- Total length of element: 650mm
- Outer Diameter: 18 MM
- Resistance of the element: 1.74 ohms
- Operation: Three phase
- Power: 9 kW line lord
- Maximum temperature: 1450 C
- Working temperature: 1400 C (For continuous operation) 4-6 hours
- Electrical line requirement: 100A - MCB (Three phase with neutral) has to be arranged by the customer
- Temperature control: Digital Temperature controller will be a PID automatic control Power control and programmable with Necessary safety features. With ramp soak
- Including 4 programming memories & atleast 8 ramp temperature and holding time.
- Temperature sensor: Pt+Rh13%/Pt - `R` type thermocouple along With Recrystallized alumina beads and Sheath
- Power control: MCB Mains
- Indications: a) Amp meter b) pilot indicating light C) Mains Indicator d) Output Indicator
- Control switches: Mains on, out put on
- Safety Control: Over heat safety cut-off of safety controller provided
- Overall Dimensions: 790 X 980 X 770 (L X W X D) mm